Complexometric & Acid–Base Titrations: Mixed Practice
Complexometric (EDTA) titration calculations mixed with acid-base titrations and general analytical practice.
Assessment Overview: 5 multiple-choice questions assessing conceptual and mathematical proficiency in Complexometric & Acid–Base Titrations: Mixed Practice.
Practice Questions & Answers
Question 1: 50 mL of tap water requires 12.36 mL of 0.01 M EDTA (Eriochrome Black T indicator) to reach the endpoint. What is the total hardness expressed as ppm CaCO3?
- A. 24.7 ppm
- B. 12.4 ppm
- C. 247.4 ppm (Correct Answer)
- D. 494.8 ppm
Pedagogical Explanation: EDTA complexes metals 1:1: mmol EDTA = 0.01 x 12.36 = 0.1236 mmol = mmol CaCO3 equivalent; mg CaCO3 = 0.1236 x 100.09 = 12.4 mg; ppm = 12.4 / 0.050 L = 247.4 ppm.
Question 2: Trimethylamine ((CH3)3N) is titrated against standard HCl. Which indicator should be used?
- A. Phenolphthalein (8.2-10.0)
- B. Methyl orange (3.1-4.4) (Correct Answer)
- C. Bromothymol blue (6.0-7.6)
- D. Methyl red (4.4-6.2)
Pedagogical Explanation: Strong acid + weak base: the equivalence pH is below 7, so an indicator changing in acidic solution is needed. Methyl orange (3.1-4.4, red to yellow) brackets the equivalence pH; phenolphthalein changes far too late (in base).
Question 3: 20.0 mL of 0.35 M HCl is titrated with 0.44 M NaOH. Volume of titrant at the equivalence point (mL)?
- A. 8
- B. 20
- C. 31.8
- D. 15.9 (Correct Answer)
Pedagogical Explanation: Stoichiometry 1:1: 1×M(analyte)×V(analyte) = 1×M(titrant)×V(titrant) → V = 15.9 mL.
Question 4: 12.0 mL of 0.25 M H2SO4 is titrated with 0.10 M NaOH. Volume of titrant at the equivalence point (mL)?
- A. 12
- B. 7.5
- C. 30
- D. 15 (Correct Answer)
Pedagogical Explanation: Stoichiometry 1:2: 1×M(analyte)×V(analyte) = 2×M(titrant)×V(titrant) → V = 15 mL.
Question 5: Error scenario: A desiccator with spent desiccant, so every crucible gains moisture This is best classified as:
- A. Random error
- B. A gross mistake
- C. Not an error at all
- D. Systematic error (Correct Answer)
Pedagogical Explanation: It biases every measurement in the same direction (affects accuracy); found by blanks, CRMs and calibration.