Detection Limits and Quantitative Precision
LOD, LOQ, replicate statistics, and where photometric error is smallest.
Assessment Overview: 5 multiple-choice questions assessing conceptual and mathematical proficiency in Detection Limits and Quantitative Precision.
Practice Questions & Answers
Question 1: Blank standard deviation σ = 0.003 AU and calibration slope m = 6400 AU·L·mol⁻¹. What is the limit of detection (LOD) in µmol·L⁻¹?
- A. 1.406 (Correct Answer)
- B. 0.4688
- C. 4.688
- D. 57.6
Pedagogical Explanation: LOD = 3σ/m = 3(0.003)/(6400) = 1.406e-06 mol·L⁻¹ = 1.406 µmol·L⁻¹.
Question 2: Study: testing two different C18 column brands. This is a test of:
- A. specificity
- B. robustness (Correct Answer)
- C. ruggedness
- D. linearity
Pedagogical Explanation: Small deliberate variations around the nominal point test robustness.
Question 3: The noise that limits LOD in trace analysis is usually
- A. room temperature
- B. long-term drift only
- C. short-term baseline noise (Correct Answer)
- D. the analyst's hearing
Pedagogical Explanation: LOD is set by the rapid baseline fluctuations around the signal.
Question 4: Control limits should be based on
- A. round numbers
- B. regulatory wishes
- C. the analyst's mood
- D. the method's own historical performance data (Correct Answer)
Pedagogical Explanation: Limits from 20+ in-control points reflect what the method actually does.
Question 5: IUPAC defines LOD as
- A. s_blank x slope
- B. 3.3 x s_blank / slope (Correct Answer)
- C. 10 x s_blank / slope
- D. 3 x mean_blank
Pedagogical Explanation: LOD = 3.3*sigma/S (sigma = blank SD, S = calibration slope).