Solvents and Sample Preparation
UV cutoffs, polarity and pH effects, and clean measurement practice.
Assessment Overview: 5 multiple-choice questions assessing conceptual and mathematical proficiency in Solvents and Sample Preparation.
Practice Questions & Answers
Question 1: Which solvent is suitable for UV measurements at 260 nm (i.e., transparent there)?
- A. Toluene
- B. Acetonitrile (Correct Answer)
- C. Dimethylformamide
- D. Carbon tetrachloride
Pedagogical Explanation: A solvent is usable only above its UV cutoff. Acetonitrile cuts off at 190 nm (< 260 nm); the others cut off at/above 260 nm.
Question 2: At 220 nm, which of these solvents will NOT interfere by absorbing?
- A. Tetrahydrofuran
- B. Toluene
- C. Acetone
- D. Hexane (Correct Answer)
Pedagogical Explanation: A solvent is usable only above its UV cutoff. Hexane cuts off at 195 nm (< 220 nm); the others cut off at/above 220 nm.
Question 3: Solids sampling: labeling each increment bag immediately. How is this practice judged?
- A. good sampling practice (Correct Answer)
- B. required by law
- C. poor sampling practice
- D. irrelevant
Pedagogical Explanation: It gives every part of the lot fair representation.
Question 4: In known-addition potentiometry, the sample concentration is found from
- A. the potential change after adding a known amount of analyte (Correct Answer)
- B. the color
- C. the temperature
- D. a single reading
Pedagogical Explanation: The ΔE from a known spike, via the Nernst equation, yields the original concentration.
Question 5: Stirring the sample during potentiometry
- A. is forbidden
- B. damages the electrode always
- C. gives uniform concentration and faster, stable readings (Correct Answer)
- D. changes the pH
Pedagogical Explanation: Gentle stirring eliminates concentration gradients at the membrane for a quicker stable potential.