CHEMVEN

Zeta Potential Analyzer

Predicts colloidal stability from surface charge

SPECTROSCOPYAnalytical Instrument

Operational Principle

Measures the zeta potential — the electrical potential at the slipping plane of a particle — which predicts whether colloids will stay suspended or aggregate. A magnitude above ~30 mV generally means a stable dispersion.

Optical & Mechanical Architecture

An electric field is applied across the sample cell and the particles' electrophoretic velocity is measured by laser Doppler velocimetry (electrophoretic light scattering): the scattered-light frequency shift Δν = 2v·sin(θ/2)/λ gives mobility μ, converted to zeta potential via the Henry equation ζ = 3ημ/(2εf(κa)). Automated titrators map zeta vs pH to find the isoelectric point where stability collapses.

Applications in Research & Industry

Suspension and emulsion formulation stability,Water-treatment coagulant dosing optimization,Nanoparticle surface-modification verification,Pigment dispersion QC in paints and inks

Typical Performance Specifications

Range ±200 mV · mobility ±10 µm·cm/V·s · size 5 nm–100 µm · pH auto-titration 1–13

Operational Visualization Notes

spectrometer

← Back to Instruments Hub