Zeta Potential Analyzer
Predicts colloidal stability from surface charge
Operational Principle
Measures the zeta potential — the electrical potential at the slipping plane of a particle — which predicts whether colloids will stay suspended or aggregate. A magnitude above ~30 mV generally means a stable dispersion.
Optical & Mechanical Architecture
An electric field is applied across the sample cell and the particles' electrophoretic velocity is measured by laser Doppler velocimetry (electrophoretic light scattering): the scattered-light frequency shift Δν = 2v·sin(θ/2)/λ gives mobility μ, converted to zeta potential via the Henry equation ζ = 3ημ/(2εf(κa)). Automated titrators map zeta vs pH to find the isoelectric point where stability collapses.
Applications in Research & Industry
Suspension and emulsion formulation stability,Water-treatment coagulant dosing optimization,Nanoparticle surface-modification verification,Pigment dispersion QC in paints and inks
Typical Performance Specifications
Operational Visualization Notes
spectrometer