WDS (Wavelength-Dispersive Spectroscopy)

High-resolution X-ray microanalysis with crystals

MICROSCOPYAnalytical Instrument

Operational Principle

Wavelength-dispersive spectroscopy is the high-precision sibling of EDX: diffracting crystals separate characteristic X-rays by wavelength, giving 10× better spectral resolution for trace and overlapping elements.

Optical & Mechanical Architecture

Characteristic X-rays from the electron-probe microanalyzer are diffracted by analyzing crystals (e.g. TAP, PET, LiF) according to Bragg's law (nλ = 2d sinθ); scanning the crystal and detector through θ selects one wavelength at a time with ~5–10 eV resolution versus ~125 eV for EDX. This resolves severe overlaps (e.g. Ti Kβ vs V Kα, S Kα vs Mo Lα) and, with high peak-to-background, reaches detection limits of ~10–100 ppm. Quantitative analysis uses full matrix corrections against standards; typical probe conditions are 15–20 kV and 10–100 nA.

Applications in Research & Industry

Trace-element quantification at 10–100 ppm levels,Resolving overlapping peaks (REE, transition metals),Precise mineral and glass stoichiometry,Diffusion profiles and zoned-crystal traverses

Typical Performance Specifications

Spectral resolution ~5–10 eV · detection limit ~10–100 ppm · crystals TAP/PET/LiF · 15–20 kV, 10–100 nA probe