CHEMVEN

Turbidimeter / Nephelometer

Quantifies cloudiness from scattered light

SPECTROSCOPYAnalytical Instrument

Operational Principle

Measures water and liquid clarity by the light scattered from suspended particles, reported in NTU (nephelometric turbidity units). It is the regulatory instrument for drinking-water safety worldwide.

Optical & Mechanical Architecture

A tungsten or 860 nm IR LED beam enters the sample; a detector at 90° measures scattered light per EPA Method 180.1 (white light) or ISO 7027 (860 nm IR, immune to sample color). The signal is calibrated against formazin polymer standards from 0.02 to 4000 NTU, with ratiometric multi-detector designs compensating for color and lamp drift. WHO requires <1 NTU for safe drinking water.

Applications in Research & Industry

Drinking-water treatment plant compliance monitoring,Wastewater discharge permitting,Beverage clarity QC (beer, wine, juices),Filtration breakthrough detection

Typical Performance Specifications

Range 0.02–4000 NTU · accuracy ±2% or ±0.01 NTU · EPA 180.1 / ISO 7027 methods · formazin calibration

Operational Visualization Notes

spectrometer

← Back to Instruments Hub