Scanning Transmission Electron Microscope (STEM)
Atomic imaging with a focused sub-ångström probe
Operational Principle
STEM rasters a sub-ångström electron probe across a thin specimen, collecting scattered electrons to image individual atomic columns with Z-contrast.
Optical & Mechanical Architecture
Electromagnetic lenses focus the beam to a probe of 0.5–1 Å (aberration-corrected), scanned pixel by pixel; an annular dark-field (ADF/HAADF) detector collects electrons scattered to high angles, where the signal scales roughly as Z^1.7–2 — heavy atoms appear bright on a dark background, giving direct chemical identification of atomic columns. Simultaneous EELS and EDX spectrum imaging maps elements and bonding at atomic resolution; bright-field and differential-phase-contrast modes image light elements and electromagnetic fields.
Applications in Research & Industry
Atomic-column imaging of catalysts and nanoparticles,Grain-boundary and interface segregation mapping,Single-atom dopant and defect identification,Atomic-resolution EELS bonding and EDS composition maps
Typical Performance Specifications
Operational Visualization Notes
microscope