Secondary Ion Mass Spectrometer (SIMS)

Traces elements and isotopes at ppb, layer by layer

SPECTROSCOPYAnalytical Instrument

Operational Principle

Blasts the surface with a primary ion beam and mass-analyzes the ejected secondary ions, delivering the most sensitive surface analysis available — parts-per-billion detection with depth profiling and isotopic information.

Optical & Mechanical Architecture

A primary beam of O₂⁺, Cs⁺ or Bi₃⁺ ions (10–30 keV) sputters the surface; secondary ions are extracted into a magnetic-sector or time-of-flight mass analyzer. Static SIMS (dose <10¹³ ions/cm²) reads the intact top monolayer; dynamic SIMS erodes continuously for depth profiles with sub-nanometer resolution. Matrix effects make quantification standard-dependent, but isotope ratios (e.g., ¹⁸O/¹⁶O) are measured with per-mil precision.

Applications in Research & Industry

Dopant depth profiles in silicon wafers,Isotopic tracing in geochemistry and cosmochemistry,Contamination analysis in semiconductor fabs,Drug distribution mapping in tissue (imaging SIMS)

Typical Performance Specifications

Primary ions O₂⁺/Cs⁺ 10–30 keV · LOD ppb range · depth resolution <1 nm · lateral resolution 50 nm (ToF) · vacuum <10⁻⁹ mbar