Phosphorimeter
Reads long-lived glow after the light is off
Operational Principle
Measures phosphorescence — the delayed emission from triplet states — to analyze compounds that fluoresce weakly or not at all. Its time-gated detection separates long-lived phosphorescence from prompt fluorescence and scatter.
Optical & Mechanical Architecture
A xenon flash lamp excites the sample; a rotating-can or electronic chopper opens the detector only after a delay of ~0.1–10 ms, when fluorescence and scatter have decayed. The remaining phosphorescence (lifetimes ms to seconds, obeying first-order decay I = I₀e^(−t/τ)) is recorded as excitation/emission spectra or decay curves. Samples are often frozen at 77 K or immobilized on filter paper to protect triplet states from oxygen quenching.
Applications in Research & Industry
Polycyclic aromatic hydrocarbon (PAH) analysis,Drug photostability and impurity profiling,Oxygen sensing via phosphorescence quenching,Forensic ink and document examination
Typical Performance Specifications