CHEMVEN

Limiting Oxygen Index Tester

How much oxygen a plastic needs to burn

THERMAL ANALYSISAnalytical Instrument

Operational Principle

Finds the minimum oxygen concentration that keeps a material burning — the single number (LOI %) behind most polymer flammability ratings.

Optical & Mechanical Architecture

A vertical sample burns in a glass chimney fed with a controlled O₂/N₂ mix. The O₂ fraction is adjusted until the sample just sustains flame for 3 minutes or 50 mm. LOI above ~27% means self-extinguishing in air; below 21% burns freely.

Applications in Research & Industry

Plastic flame-retardant grading,Textile flammability,Cable insulation specs,Foam safety testing

Typical Performance Specifications

O₂/N₂ mixing column · LOI % readout · ASTM D2863

Operational Visualization Notes

furnace

← Back to Instruments Hub