Kelvin Probe Force Microscope (KPFM)

Nanoscale work-function and surface-potential maps

MICROSCOPYAnalytical Instrument

Operational Principle

Kelvin probe force microscopy maps the contact potential difference — and hence work function — of surfaces at nanometer resolution using a conductive AFM tip.

Optical & Mechanical Architecture

A conductive tip (Pt/Ir or doped diamond) hovers over the sample while an AC voltage plus a DC bias is applied; the electrostatic force between tip and sample oscillates at the AC frequency, and a feedback loop adjusts the DC bias to null it. That nulling voltage equals the contact potential difference (CPD) between tip and sample, mapping local work function with ~10–50 nm lateral resolution and millivolt sensitivity. FM-KPFM detects the force gradient for sharper resolution; measurements work in air, vacuum or liquid.

Applications in Research & Industry

Work-function mapping of semiconductors and 2D materials,Corrosion initiation and coating-defect potential imaging,Organic solar-cell and perovskite band-alignment studies,Triboelectric and charge-trapping surface analysis

Typical Performance Specifications

Lateral ~10–50 nm · voltage sensitivity ~1–10 mV · conductive tips (Pt/Ir) · AM or FM detection · air/vacuum/liquid operation