Differential Mobility Analyzer (DMA)
Selects nanoparticles by electrical mobility
Operational Principle
A differential mobility analyzer that classifies aerosol nanoparticles by size with nanometer resolution — the heart of scanning mobility particle sizers used in air-quality and materials labs.
Optical & Mechanical Architecture
Charged particles enter a laminar sheath-air flow between two electrodes; the electric field deflects them by electrical mobility, and only one narrow mobility band exits through a slit. Scanning the voltage with a condensation particle counter downstream builds a full size distribution.
Applications in Research & Industry
Air-quality ultrafine particle monitoring,Engineered nanoparticle size QC,Atmospheric nucleation studies,Filter and respirator efficiency testing
Typical Performance Specifications
3–800 nm · sheath flow 1–20 L/min · scanned with CPC detector
Operational Visualization Notes
beam