Capillary Gel Electrophoresis
Automated high-resolution fragment sizing
Operational Principle
Replaces slab gels with a polymer-filled capillary — automated, quantitative DNA/protein sizing in minutes.
Optical & Mechanical Architecture
Sample is electrokinetically injected into a narrow capillary filled with sieving polymer (e.g. linear polyacrylamide); fragments separate by size under high field and pass a fluorescence detector. Heat dissipates fast, so voltages reach ~15–30 kV.
Applications in Research & Industry
Sanger sequencing prep,Fragment analysis (microsatellites),Oligo purity,Protein size distribution
Typical Performance Specifications
Capillary 50 µm id, 30–50 cm · field 200–600 V/cm · bp resolution 1–2 bp
Operational Visualization Notes