Biosafety Cabinet

Sterile work zone that protects the user

GENERAL LABORATORYAnalytical Instrument

Operational Principle

A ventilated enclosure that protects the operator, the sample and the environment when working with infectious agents — airflow in, HEPA-filtered air down, exhaust out.

Optical & Mechanical Architecture

Room air is drawn in at the front grille (personnel protection); HEPA-filtered laminar downflow bathes the work surface (product protection); exhaust passes through a second HEPA filter before leaving (environmental protection). Class II Type A2 is the common lab standard; the sash must stay at the marked height for the air curtain to work.

Applications in Research & Industry

Microbiological culture work,Clinical sample handling,Cell culture with pathogens,Sterile media preparation

Typical Performance Specifications

Class II Type A2 · inflow ≥0.5 m/s · downflow ~0.3 m/s · 99.995% HEPA at 0.3 µm