Auger Electron Spectrometer (AES)
Nanometer-scale surface elemental mapping
Operational Principle
Maps surface elemental composition with nanometer lateral resolution by detecting Auger electrons — a three-electron process that, like XPS, fingerprints every element except hydrogen and helium in the top few nanometers.
Optical & Mechanical Architecture
A focused electron beam (3–30 keV) ejects a core electron; a higher electron drops in and the released energy ejects a second (Auger) electron with element-specific kinetic energy, e.g., the KLL transition. A cylindrical mirror analyzer scans dN/dE, and the beam is rastered to build elemental maps with ~10 nm resolution — far better than XPS. UHV and charge compensation are used for insulators.
Applications in Research & Industry
Semiconductor failure analysis and dopant profiling,Grain-boundary segregation in metallurgy,Oxide thickness on microelectronics,Adhesion-failure surface chemistry
Typical Performance Specifications
Operational Visualization Notes
beam