Heat Capacity of the Calorimeter by Electrical Calibration
To determine the total heat capacity of a calorimeter by electrical heating (V·I·t method).
Theoretical Foundations
Passing current I at voltage V for time t dissipates q = VIt joules into the calorimeter. With water of known mass present, (m_w·c_w + C_cal)ΔT = VIt, giving C_cal directly. Electrical calibration is the most accurate route to the calorimeter constant because the energy input is measured electrically, not thermally.
Required Apparatus & Chemicals
Laboratory Apparatus
- Calorimeter with immersion heater
- DC power supply
- Voltmeter
- Ammeter
- Stopwatch
- Thermometer (0.1 °C)
- Balance
Chemical Reagents
- Distilled water
Step-by-Step Practical Procedure
- Place 200.0 g water in the calorimeter; record T.
- Set ~6.0 V across the heater; record V and I.
- Switch on for exactly 120 s with stirring.
- Record the temperature rise ΔT.
- C_cal = (VIt/ΔT) - m_w·c_w.
- Use this C_cal to correct a neutralisation result.
Observations & Expected Results
Typical student data: V = 6.0 V, I = 2.0 A, t = 120 s, ΔT = 1.60 °C with 200.0 g water → C_cal ≈ 64 J/°C for a glass calorimeter (larger than a polystyrene cup).
Worked Stoichiometric Calculations
Problem: Typical student data: V = 6.0 V, I = 2.0 A, t = 120 s, 200.0 g water, ΔT = 1.60 °C. Find C_cal.
Working: q = VIt = 6.0 × 2.0 × 120 = 1440 J. Total heat capacity = 1440/1.60 = 900 J/°C. Water contributes 200.0 × 4.18 = 836 J/°C, so C_cal = 900 - 836 = 64 J/°C. (Sanity check: C_total must exceed the water contribution — a useful trap for sign/unit errors.)
Safety Precautions & Hazard Control
- Supervision: electrical heating near water — check insulation and use low voltage.
- Do not run the heater dry.
- Stir to avoid local boiling at the element.
Waste Disposal Protocol
Water to drain.
Viva Voce Oral Defense Questions & Answers
Q: Why is electrical calibration more accurate than the mixing method?
A: Electrical energy VIt is measured with high precision; no hot-water transfer losses occur.
Q: What does C_cal include?
A: The heat capacity of the vessel, lid, stirrer, thermometer and heater — everything that warms with the water.
Q: The worked example first gave an impossible result — what was wrong?
A: The stated ΔT was inconsistent with the energy input; checking that q/ΔT exceeds the water's own heat capacity is an essential sanity test.