Relative Viscosity by the Ostwald Viscometer
To determine the relative and absolute viscosity of ethanol using an Ostwald viscometer.
Theoretical Foundations
For the same viscometer, Poiseuille's law gives eta1/eta2 = (rho1·t1)/(rho2·t2), where t is the flow time between the marks. Measuring against water (eta = 1.00 mPa·s at 20 C) yields the unknown viscosity.
Required Apparatus & Chemicals
Laboratory Apparatus
- Ostwald viscometer
- Thermostat/water bath
- Stopwatch
- Pipette
- Density bottle
- Clamp and stand
Chemical Reagents
- Distilled water (reference)
- Ethanol
- Acetone (for drying)
- Chromic acid (cleaning)
Step-by-Step Practical Procedure
- Clean and dry the viscometer; clamp it vertically in the thermostat at 25 C.
- Pipette a fixed volume of distilled water; suck it above the upper mark.
- Measure the flow time t1 between marks (3 repeats).
- Rinse, dry, repeat with ethanol to get t2.
- Determine densities; calculate eta_rel = (t2·rho2)/(t1·rho1).
- Multiply by eta_water to get absolute viscosity.
Observations & Expected Results
Typical student data: water t1 = 120.0 s, ethanol t2 = 182 s; eta_rel ~ 1.20, giving eta ~ 1.20 mPa·s (literature 1.20 at 20 C).
Worked Stoichiometric Calculations
Problem: t_water = 120 s, rho_water = 0.998; t_ethanol = 182 s, rho_ethanol = 0.789. eta_water = 1.00 mPa·s. Find eta_ethanol.
Working: eta_rel = (182 x 0.789)/(120 x 0.998) = 143.6/119.8 = 1.199. eta_ethanol = 1.199 x 1.00 = 1.20 mPa·s.
Safety Precautions & Hazard Control
- The viscometer must be exactly vertical.
- Use identical liquid volumes each run.
- Keep temperature within +-0.1 C (viscosity is very T-sensitive).
- Never blow by mouth - use a pipette filler.
Waste Disposal Protocol
Ethanol to organic waste; rinse water to drain.
Viva Voce Oral Defense Questions & Answers
Q: Why is temperature control critical?
A: Viscosity falls ~2% per C for most liquids; a 1 C drift ruins the comparison.
Q: What is the kinetic-energy (Hagenbach) correction?
A: A small correction for the kinetic energy of the emerging jet, significant for short flow times.
Q: Why use the same volume each time?
A: The driving pressure head depends on the liquid levels; equal volumes keep the geometry identical.