Fluoride Determination with an Ion-Selective Electrode
To determine fluoride in a sample using a fluoride ion-selective electrode.
Theoretical Foundations
The LaF3 crystal membrane electrode responds as E = const - 0.0591 log[F-] at 25 C (Nernstian, one decade = 59 mV). TISAB buffer fixes ionic strength and pH and complexes interfering Al3+/Fe3+.
Required Apparatus & Chemicals
Laboratory Apparatus
- Fluoride ISE
- Calomel or Ag/AgCl reference electrode
- Digital millivoltmeter
- Beakers
- Volumetric flasks
Chemical Reagents
- Sodium fluoride standard (1000 ppm)
- TISAB buffer
- Unknown fluoride sample (e.g. toothpaste extract)
Step-by-Step Practical Procedure
- Prepare F- standards of 1, 10 and 100 ppm, each 1:1 with TISAB.
- Measure E for each standard.
- Plot E vs log[F-]; check slope ~ -59 mV/decade.
- Measure the unknown (with TISAB) and read its concentration from the plot.
Observations & Expected Results
Slope ~ -57 to -59 mV/decade; unknown reads within the calibrated range.
Worked Stoichiometric Calculations
Problem: Standards give E = -118 mV at 10 ppm and -177 mV at 100 ppm. The unknown gives -148 mV. Find [F-].
Working: Slope = (-177 + 118)/1 = -59 mV/decade. log[F-] = log(10) + (-148 + 118)/(-59) = 1 + 30/59 = 1 + 0.508 = 1.508. [F-] = 10^1.508 = 32.2 ppm, about 32 ppm.
Safety Precautions & Hazard Control
- Always add TISAB to standards and samples alike.
- Rinse electrodes between readings.
- Switch off the power supply before changing any connections.
Waste Disposal Protocol
Fluoride solutions to aqueous chemical waste.
Viva Voce Oral Defense Questions & Answers
Q: Why is TISAB essential?
A: It holds ionic strength and pH constant (so activity ~ concentration) and frees F- from Al3+/Fe3+ complexes.
Q: Why is the slope negative?
A: F- is an anion: E = const - 0.0591 log[F-], so higher [F-] gives a more negative potential.