Calibration of the Glass Electrode with Standard Buffers

To calibrate a glass pH electrode and check its Nernstian slope.

Level: UniversitySafety: SafeTime: 45 min
Aim of Experiment: To calibrate a glass pH electrode and check its Nernstian slope.

Theoretical Foundations

The glass electrode responds as E = const - 0.0591 pH at 25 C. Two buffers (pH 4.00 and 7.00, or 7.00 and 10.00) fix the constant and the slope; a good electrode gives 57-59 mV/pH.

Required Apparatus & Chemicals

Laboratory Apparatus

  • pH meter with glass electrode
  • Beakers
  • Wash bottle
  • Thermometer

Chemical Reagents

  • Standard buffer solutions pH 4.00, 7.00 and 10.00
  • Distilled water (for rinsing the electrode)
  • Electrode storage solution (3 M KCl)

Step-by-Step Practical Procedure

  1. Rinse the electrode and immerse in pH 7.00 buffer; set the meter to 7.00.
  2. Rinse and immerse in pH 4.00 buffer; check the slope reading (should be ~100%).
  3. Record the mV readings in each buffer.
  4. Calculate mV/pH from the difference.
  5. Test the pH 10.00 buffer as a check.

Observations & Expected Results

mV difference between pH 7 and pH 4 ~ 177 mV, i.e. 59 mV/pH; slope 98-102%.

Worked Stoichiometric Calculations

Problem: The meter reads +177 mV in pH 4.00 buffer and 0 mV in pH 7.00 buffer. Find the slope in mV/pH.

Working: Slope = 177 mV/3 pH units = 59.0 mV/pH - Nernstian response.

Safety Precautions & Hazard Control

  • Never wipe the glass bulb; rinse only.
  • Keep the bulb hydrated in storage solution.
  • Wear safety goggles and a lab coat throughout the experiment.

Waste Disposal Protocol

Buffers can go down the drain.

Viva Voce Oral Defense Questions & Answers

Q: Why calibrate with two buffers?

A: One fixes the offset (asymmetry potential) and the second fixes the slope; a single point cannot do both.

Q: What does a slope below 55 mV/pH indicate?

A: An aged or dehydrated glass membrane - the electrode needs rejuvenation or replacement.