Calibration of the Glass Electrode with Standard Buffers
To calibrate a glass pH electrode and check its Nernstian slope.
Theoretical Foundations
The glass electrode responds as E = const - 0.0591 pH at 25 C. Two buffers (pH 4.00 and 7.00, or 7.00 and 10.00) fix the constant and the slope; a good electrode gives 57-59 mV/pH.
Required Apparatus & Chemicals
Laboratory Apparatus
- pH meter with glass electrode
- Beakers
- Wash bottle
- Thermometer
Chemical Reagents
- Standard buffer solutions pH 4.00, 7.00 and 10.00
- Distilled water (for rinsing the electrode)
- Electrode storage solution (3 M KCl)
Step-by-Step Practical Procedure
- Rinse the electrode and immerse in pH 7.00 buffer; set the meter to 7.00.
- Rinse and immerse in pH 4.00 buffer; check the slope reading (should be ~100%).
- Record the mV readings in each buffer.
- Calculate mV/pH from the difference.
- Test the pH 10.00 buffer as a check.
Observations & Expected Results
mV difference between pH 7 and pH 4 ~ 177 mV, i.e. 59 mV/pH; slope 98-102%.
Worked Stoichiometric Calculations
Problem: The meter reads +177 mV in pH 4.00 buffer and 0 mV in pH 7.00 buffer. Find the slope in mV/pH.
Working: Slope = 177 mV/3 pH units = 59.0 mV/pH - Nernstian response.
Safety Precautions & Hazard Control
- Never wipe the glass bulb; rinse only.
- Keep the bulb hydrated in storage solution.
- Wear safety goggles and a lab coat throughout the experiment.
Waste Disposal Protocol
Buffers can go down the drain.
Viva Voce Oral Defense Questions & Answers
Q: Why calibrate with two buffers?
A: One fixes the offset (asymmetry potential) and the second fixes the slope; a single point cannot do both.
Q: What does a slope below 55 mV/pH indicate?
A: An aged or dehydrated glass membrane - the electrode needs rejuvenation or replacement.