Potentiometric Argentometric Titration: Chloride with AgNO3

To determine chloride by potentiometric titration with AgNO3 using a silver electrode.

Level: UniversitySafety: Needs supervisionTime: 60 min
Aim of Experiment: To determine chloride by potentiometric titration with AgNO3 using a silver electrode.

Theoretical Foundations

Ag+ + Cl- -> AgCl(s). A silver indicator electrode follows E = Edeg + 0.0591 log[Ag+]. Before the endpoint [Ag+] is tiny (fixed by Ksp); after, it rises sharply, giving a large potential jump at the endpoint located from DeltaE/DeltaV.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Silver wire indicator electrode
  • Calomel reference electrode
  • Digital voltmeter
  • Burette
  • Beaker
  • Stirrer

Chemical Reagents

  • Sodium chloride (unknown, ~0.05 M)
  • Silver nitrate (0.050 M standard)
  • Distilled water (for rinsing glassware and electrodes)

Step-by-Step Practical Procedure

  1. Pipette 25.0 mL of the chloride sample.
  2. Insert the Ag and reference electrodes.
  3. Titrate with 0.050 M AgNO3, recording E in 0.5 mL steps (0.1 mL near the jump).
  4. Plot E and DeltaE/DeltaV vs V.
  5. Find the endpoint and calculate [Cl-].

Observations & Expected Results

E ~ 0.3 V initially, jumping ~200 mV at the endpoint.

Worked Stoichiometric Calculations

Problem: Endpoint at 24.4 mL of 0.0500 M AgNO3 for 25.0 mL sample. Find [Cl-].

Working: 1:1 reaction. Moles Ag+ = 0.0500 x 0.0244 = 0.00122 mol = moles Cl-. [Cl-] = 0.00122/0.0250 = 0.0488 M.

Safety Precautions & Hazard Control

  • AgNO3 stains; wear gloves and protect from light.
  • Clean the Ag electrode before use.
  • Switch off the power supply before changing any connections.

Waste Disposal Protocol

Collect AgCl precipitate and solutions for silver recovery.

Viva Voce Oral Defense Questions & Answers

Q: Why is there a sudden potential jump at the endpoint?

A: [Ag+] is pinned at ~10^-5 M by Ksp before the endpoint but rises freely with excess Ag+ after it - a huge logarithmic change.

Q: Why is no indicator needed?

A: The silver electrode itself signals the endpoint through the potential jump.