pH Determination with the Quinhydrone Electrode

To measure pH using a quinhydrone electrode and verify its Nernstian response.

Level: UniversitySafety: Needs supervisionTime: 60 min
Aim of Experiment: To measure pH using a quinhydrone electrode and verify its Nernstian response.

Theoretical Foundations

Quinhydrone (quinone/hydroquinone 1:1) on Pt gives: Q + 2H+ + 2e- <=> H2Q, so E = Edeg - 0.0591 pH with Edeg = +0.699 V at 25 C. Measuring E against a reference gives pH = (0.699 - E)/0.0591.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Platinum wire electrode
  • Calomel reference electrode
  • Digital voltmeter
  • Beakers
  • Salt bridge

Chemical Reagents

  • Quinhydrone (solid)
  • Buffer solutions of known pH
  • Test solution (unknown pH)

Step-by-Step Practical Procedure

  1. Add a pinch of quinhydrone to each buffer and stir.
  2. Measure E vs the calomel electrode for pH 4, 7 and 9 buffers.
  3. Plot E vs pH; check slope ~ -59 mV/pH.
  4. Measure the unknown and compute its pH.
  5. Compare with a glass-electrode reading.

Observations & Expected Results

Slope ~ -58 to -60 mV/pH; unknown pH agrees within 0.05 units.

Worked Stoichiometric Calculations

Problem: E measured = 0.287 V vs SHE at 25 C. Calculate the pH.

Working: pH = (0.699 - 0.287)/0.0591 = 0.412/0.0591 = 6.97, about pH 7.0.

Safety Precautions & Hazard Control

  • Quinhydrone is toxic; wear gloves.
  • Use fresh solutions; quinhydrone oxidises in air.
  • Switch off the power supply before changing any connections.

Waste Disposal Protocol

Collect quinhydrone waste as organic chemical waste.

Viva Voce Oral Defense Questions & Answers

Q: Why does the quinhydrone electrode fail above pH 8?

A: Hydroquinone is a weak acid and dissociates, and quinone reacts with OH-, destroying the 1:1 ratio.

Q: What is the advantage over the hydrogen electrode?

A: No gas supply is needed; it is a simple dip electrode - but limited to acidic/neutral solutions.