pH Determination with the Quinhydrone Electrode
To measure pH using a quinhydrone electrode and verify its Nernstian response.
Theoretical Foundations
Quinhydrone (quinone/hydroquinone 1:1) on Pt gives: Q + 2H+ + 2e- <=> H2Q, so E = Edeg - 0.0591 pH with Edeg = +0.699 V at 25 C. Measuring E against a reference gives pH = (0.699 - E)/0.0591.
Required Apparatus & Chemicals
Laboratory Apparatus
- Platinum wire electrode
- Calomel reference electrode
- Digital voltmeter
- Beakers
- Salt bridge
Chemical Reagents
- Quinhydrone (solid)
- Buffer solutions of known pH
- Test solution (unknown pH)
Step-by-Step Practical Procedure
- Add a pinch of quinhydrone to each buffer and stir.
- Measure E vs the calomel electrode for pH 4, 7 and 9 buffers.
- Plot E vs pH; check slope ~ -59 mV/pH.
- Measure the unknown and compute its pH.
- Compare with a glass-electrode reading.
Observations & Expected Results
Slope ~ -58 to -60 mV/pH; unknown pH agrees within 0.05 units.
Worked Stoichiometric Calculations
Problem: E measured = 0.287 V vs SHE at 25 C. Calculate the pH.
Working: pH = (0.699 - 0.287)/0.0591 = 0.412/0.0591 = 6.97, about pH 7.0.
Safety Precautions & Hazard Control
- Quinhydrone is toxic; wear gloves.
- Use fresh solutions; quinhydrone oxidises in air.
- Switch off the power supply before changing any connections.
Waste Disposal Protocol
Collect quinhydrone waste as organic chemical waste.
Viva Voce Oral Defense Questions & Answers
Q: Why does the quinhydrone electrode fail above pH 8?
A: Hydroquinone is a weak acid and dissociates, and quinone reacts with OH-, destroying the 1:1 ratio.
Q: What is the advantage over the hydrogen electrode?
A: No gas supply is needed; it is a simple dip electrode - but limited to acidic/neutral solutions.