CHEMVEN

Potentiometric Redox Titration: Fe2+ versus Cerium(IV)

To determine Fe2+ by potentiometric titration with Ce(SO4)2.

Level: UniversitySafety: Needs supervisionTime: 75 min
Aim of Experiment: To determine Fe2+ by potentiometric titration with Ce(SO4)2.

Theoretical Foundations

Ce4+ + Fe2+ -> Ce3+ + Fe3+. A Pt electrode follows E = Edeg + 0.0591 log([ox]/[red]). Before the endpoint the Fe3+/Fe2+ couple (0.77 V) controls E; after, the Ce4+/Ce3+ couple (1.44 V) takes over, giving a large potential jump at the endpoint.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Pt indicator electrode
  • Calomel reference electrode
  • Digital voltmeter
  • Burette
  • Beaker
  • Stirrer

Chemical Reagents

  • Ferrous ammonium sulphate (~0.05 M in 1 M H2SO4)
  • Cerium(IV) sulphate (0.050 M standard)
  • Distilled water (for rinsing glassware and electrodes)

Step-by-Step Practical Procedure

  1. Pipette 25.0 mL of the Fe2+ solution.
  2. Insert Pt and reference electrodes.
  3. Add Ce4+ in 1.0 mL steps (0.1 mL near the jump), recording E.
  4. Plot E vs V and DeltaE/DeltaV vs V.
  5. Find the endpoint and calculate [Fe2+].

Observations & Expected Results

E ~ 0.7 V initially, jumping to ~1.3 V at the endpoint within 0.2 mL.

Worked Stoichiometric Calculations

Problem: Endpoint at 23.6 mL of 0.0500 M Ce4+ for 25.0 mL Fe2+ solution. Find [Fe2+].

Working: Reaction is 1:1. Moles Ce4+ = 0.0500 x 0.0236 = 0.00118 mol = moles Fe2+. [Fe2+] = 0.00118/0.0250 = 0.0472 M.

Safety Precautions & Hazard Control

  • Keep the medium strongly acidic to stabilise Ce4+.
  • Clean the Pt electrode before use.
  • Switch off the power supply before changing any connections.

Waste Disposal Protocol

Collect cerium/iron waste in the heavy-metal bottle.

Viva Voce Oral Defense Questions & Answers

Q: What is the potential exactly at the equivalence point?

A: For a 1:1 electron stoichiometry, Eeq = (E1 + E2)/2 = (0.77 + 1.44)/2 = 1.105 V.

Q: Why is Ce4+ preferred over KMnO4 for potentiometry?

A: Ce4+ gives a single clean 1-electron step and stable standard solutions, while MnO4- chemistry is more complex.