Temperature Dependence of Electrolytic Conductivity

To measure the temperature coefficient of conductivity of an electrolyte.

Level: UniversitySafety: SafeTime: 45 min
Aim of Experiment: To measure the temperature coefficient of conductivity of an electrolyte.

Theoretical Foundations

Conductivity rises ~2% per C because viscosity falls and ionic mobilities increase. The temperature coefficient alpha = (1/kappa)(dkappa/dT) is about 0.02 K^-1 for most salts. This is why conductivity work needs thermostats.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Conductivity cell
  • Conductivity meter
  • Water baths (20-50 C)
  • Thermometer

Chemical Reagents

  • Sodium chloride (0.010 M)
  • Distilled water (for dilutions)
  • Ice (for the low-temperature bath)

Step-by-Step Practical Procedure

  1. Thermostat the cell at 20 C and record kappa.
  2. Repeat at 30, 40 and 50 C.
  3. Plot kappa vs T.
  4. Calculate alpha from the slope.
  5. Compare with the typical 2% per C.

Observations & Expected Results

kappa rises linearly; typical alpha ~ 0.021 K^-1.

Worked Stoichiometric Calculations

Problem: kappa = 1185 microS/cm at 20 C and 1440 microS/cm at 40 C. Find the mean temperature coefficient.

Working: alpha = (1440 - 1185)/(1185 x 20) = 255/23,700 = 0.0108 per C, about 1.1% per C (illustrative; typical lab values 1.5-2.5% per C).

Safety Precautions & Hazard Control

  • Allow full thermal equilibration at each temperature.
  • Do not overheat the cell.
  • Wear safety goggles and a lab coat throughout the experiment.

Waste Disposal Protocol

NaCl solution can go down the drain.

Viva Voce Oral Defense Questions & Answers

Q: Why does conductivity increase with temperature?

A: Lower water viscosity lets ions move faster; also, ion-pair dissociation increases slightly.

Q: Why is temperature control critical in conductometry?

A: A 1 C drift changes kappa by ~2%, which would swamp small analytical signals.