Conductometric Titration of a Strong and Weak Acid Mixture

To determine both acids in an HCl + CH3COOH mixture by a single conductometric titration.

Level: UniversitySafety: Needs supervisionTime: 75 min
Aim of Experiment: To determine both acids in an HCl + CH3COOH mixture by a single conductometric titration.

Theoretical Foundations

The strong acid is neutralised first: conductance falls (H+ replaced by Na+) to a first minimum (V1). Then the weak acid is neutralised: conductance rises gradually as CH3COONa forms (V2). Excess NaOH gives a steep rise. V1 gives HCl; V2 - V1 gives CH3COOH.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Conductivity meter
  • Burette
  • Beaker
  • Magnetic stirrer

Chemical Reagents

  • Mixture of HCl and acetic acid (unknown)
  • Sodium hydroxide (0.10 M standard)
  • Distilled water (for rinsing glassware and electrodes)

Step-by-Step Practical Procedure

  1. Pipette 25.0 mL of the mixture, dilute to ~100 mL.
  2. Titrate with 0.10 M NaOH in 1.0 mL steps, recording conductance.
  3. Plot conductance vs volume.
  4. Locate the first minimum (V1) and the second break (V2).
  5. Calculate each acid's concentration.

Observations & Expected Results

Curve falls to a minimum, rises with a moderate slope, then rises steeply - two clear endpoints.

Worked Stoichiometric Calculations

Problem: V1 = 12.5 mL and V2 = 32.5 mL of 0.100 M NaOH for 25.0 mL of mixture. Find [HCl] and [CH3COOH].

Working: Moles HCl = 0.100 x 0.0125 = 0.00125 mol; [HCl] = 0.00125/0.0250 = 0.0500 M. Moles CH3COOH = 0.100 x (0.0325 - 0.0125) = 0.00200 mol; [CH3COOH] = 0.00200/0.0250 = 0.0800 M.

Safety Precautions & Hazard Control

  • Add titrant slowly near the expected breaks.
  • Maintain constant temperature.
  • Wear safety goggles and a lab coat throughout the experiment.

Waste Disposal Protocol

Neutralised solution can go down the drain.

Viva Voce Oral Defense Questions & Answers

Q: Why is the strong acid neutralised first?

A: H+ from the strong acid is present at far higher concentration; the weak acid stays largely undissociated until the strong acid is consumed.

Q: Why can't an indicator easily give both endpoints?

A: The first endpoint lies in acidic solution where most indicators do not change sharply; conductivity shows both breaks clearly.