Standard Addition Potentiometry: Determining Unknown Cu2+ Concentration

To find an unknown Cu2+ concentration by the standard addition method.

Level: UniversitySafety: Needs supervisionTime: 60 min
Aim of Experiment: To find an unknown Cu2+ concentration by the standard addition method.

Theoretical Foundations

In standard addition, known increments of standard Cu2+ are added to the unknown and E is measured after each. Since E = const + (0.0591/2) log c_total, plotting 10^(2E/0.0591) versus added concentration gives a line whose x-intercept is the unknown concentration.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Cu indicator electrode
  • Calomel reference electrode
  • Digital voltmeter
  • Beakers
  • Micropipette/burette

Chemical Reagents

  • Unknown CuSO4 solution
  • CuSO4 standard (0.10 M)
  • KNO3 (ionic strength buffer)

Step-by-Step Practical Procedure

  1. Place 50.0 mL of the unknown in a beaker with the electrodes.
  2. Record E0.
  3. Add 1.00 mL of 0.10 M CuSO4 standard, stir, record E.
  4. Repeat additions up to 5 mL.
  5. Plot 10^(2E/0.0591) vs added concentration; extrapolate to find the unknown.

Observations & Expected Results

E rises stepwise by ~5-10 mV per addition; the linearised plot extrapolates to the true unknown concentration within a few percent.

Worked Stoichiometric Calculations

Problem: After additions, the linearised plot gives an x-intercept of -0.024 M. What was the unknown concentration (50.0 mL sample)?

Working: The intercept magnitude is the concentration added-equivalent: c_unknown x (50.0 mL) corresponds to 0.024 M in the final volume. Approximately, c_unknown ~ 0.024 M (exact treatment corrects for dilution: ~0.026 M).

Safety Precautions & Hazard Control

  • Keep ionic strength constant with KNO3.
  • Allow E to stabilise after each addition.
  • Switch off the power supply before changing any connections.

Waste Disposal Protocol

Collect copper solutions in the heavy-metal waste container.

Viva Voce Oral Defense Questions & Answers

Q: What is the advantage of standard addition over direct calibration?

A: It compensates for matrix effects, since the standard experiences the same solution conditions as the unknown.

Q: Why is the plot linearised with 10^(2E/0.0591)?

A: Because E is logarithmic in concentration; the antilog transform restores a linear dependence on total concentration.