Silver-Silver Chloride Reference Electrode
To prepare an Ag/AgCl reference electrode and verify its standard potential.
Theoretical Foundations
The electrode is Ag(s)|AgCl(s)|Cl- with AgCl + e- -> Ag + Cl-, Edeg = +0.222 V. The potential depends on Cl- activity: E = 0.222 - 0.0591 log[Cl-]. In saturated KCl it is +0.197 V vs SHE at 25 C.
Required Apparatus & Chemicals
Laboratory Apparatus
- Silver wire
- DC power supply
- Beaker
- Digital voltmeter
- Salt bridge
- Cu2+/Cu half-cell
Chemical Reagents
- Hydrochloric acid (0.1 M, for chloridising)
- Potassium chloride (saturated)
- CuSO4 (1.0 M)
Step-by-Step Practical Procedure
- Clean the silver wire and make it the anode in 0.1 M HCl at ~1 mA for 10 minutes to coat it with AgCl.
- Rinse and place the wire in saturated KCl.
- Couple against Cu2+/Cu via a salt bridge and measure the EMF.
- Calculate E(Ag/AgCl) = 0.34 - Ecell; compare with 0.197 V.
Observations & Expected Results
The wire gains a uniform mauve-grey AgCl coating. Typical Ecell ~ 0.143 V, giving E(Ag/AgCl) ~ +0.197 V.
Worked Stoichiometric Calculations
Problem: An Ag/AgCl electrode in 1.0 M KCl is used. Calculate its potential vs SHE at 25 C.
Working: E = Edeg - 0.0591 log[Cl-] = 0.222 - 0.0591 log(1.0) = 0.222 - 0 = +0.222 V vs SHE.
Safety Precautions & Hazard Control
- Keep the AgCl coating in the dark; it is light-sensitive.
- Do not scratch off the coating.
- Switch off the power supply before changing any connections.
Waste Disposal Protocol
Collect silver-containing waste separately for silver recovery.
Viva Voce Oral Defense Questions & Answers
Q: Why is the Ag/AgCl electrode preferred over calomel today?
A: It contains no toxic mercury and works at higher temperatures where calomel disproportionates.
Q: What happens to its potential if [Cl-] is halved?
A: E rises by 0.0591 log 2 = 0.0178 V, about 18 mV, per the Nernst equation.