Beer-Lambert Law: Quantitative UV Analysis by Calibration Curve
To determine the concentration of an unknown solution using a UV calibration curve.
Theoretical Foundations
Beer-Lambert law: A = epsilon.c.l. Absorbance at lambda-max is plotted versus known concentrations to give a straight line through the origin; the unknown's absorbance is read off the curve. Deviations occur at high concentration or with stray light.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-visible spectrophotometer
- Cuvettes (1 cm)
- Volumetric flasks
- Pipettes
Chemical Reagents
- Potassium permanganate or methylene blue standard solutions
- Unknown solution of the same solute
- Distilled water (blank)
Step-by-Step Practical Procedure
- Prepare five standard solutions by serial dilution.
- Zero the instrument with the water blank at lambda-max.
- Measure absorbance of each standard.
- Plot A versus concentration; check linearity.
- Measure the unknown and read its concentration from the curve.
Observations & Expected Results
Linear plot through the origin (e.g. A = 0.120 at 1.0 x 10^-5 M). Unknown A = 0.600 gives c = 5.0 x 10^-5 M.
Worked Stoichiometric Calculations
Problem: A = 0.600, epsilon = 12000 L/mol/cm, l = 1 cm. Calculate c, and check it lies on the calibration line.
Working: c = A/(epsilon.l) = 0.600/12000 = 5.0 x 10^-5 mol/L. It must fall within the linear range of the plotted standards; if the curve bends here, dilute the unknown.
Safety Precautions & Hazard Control
- Wipe cuvettes clean; fingerprints absorb.
- Keep absorbance below ~1.0 for linearity.
- Use matched cuvettes and the same wavelength throughout.
Waste Disposal Protocol
Dilute dye solutions to aqueous waste.
Viva Voce Oral Defense Questions & Answers
Q: State the Beer-Lambert law and its limits.
A: A = epsilon.c.l; valid for dilute solutions, monochromatic light, and non-associating solutes; fails at high concentration or with chemical equilibria.
Q: Why must absorbance stay below ~1?
A: Above A = 1 less than 10% of light is transmitted; stray light and detector noise cause negative deviation from linearity.
Q: Why use lambda-max?
A: Sensitivity is highest (largest epsilon) and small wavelength errors matter least at the peak.