Iodine Value of an Oil: Degree of Unsaturation
To determine the iodine value of an oil by the Wijs method.
Theoretical Foundations
Iodine value = grams of I2 absorbed per 100 g of oil, measuring C=C unsaturation: -CH=CH- + ICl -> -CHI-CHCl-. Excess Wijs reagent (ICl in acetic acid) is treated with KI, liberating I2 titrated with Na2S2O3. IV = (V_blank - V_sample) x N x 12.69 / w. Drying oils have IV > 130.
Required Apparatus & Chemicals
Laboratory Apparatus
- Iodine flasks (stoppered)
- Burette
- Pipette
- Dark cupboard
- Weighing bottle
Chemical Reagents
- Oil sample (~0.5 g)
- Wijs reagent (ICl in glacial acetic acid)
- 10% KI solution
- 0.1 N Na2S2O3
- Starch indicator
- Carbon tetrachloride or chloroform
Step-by-Step Practical Procedure
- Weigh ~0.5 g oil into an iodine flask; dissolve in 10 mL CCl4.
- Add 25 mL Wijs reagent, stopper, and keep in the dark for 30 min.
- Add 10 mL KI solution and 50 mL water.
- Titrate liberated I2 with 0.1 N Na2S2O3 to pale yellow, then starch to colourless.
- Run a blank and calculate the iodine value.
Observations & Expected Results
Starch end point: blue to colourless. Typical IV: coconut ~10, olive ~85, sunflower ~130, linseed ~180 g I2/100 g.
Worked Stoichiometric Calculations
Problem: 0.500 g oil: blank 40.0 mL, sample 15.0 mL of 0.100 N Na2S2O3. Calculate the iodine value.
Working: IV = (40.0 - 15.0) x 0.100 x 12.69 / 0.500 = 25.0 x 0.100 x 12.69 / 0.500 = 63.45 g I2 per 100 g oil.
Safety Precautions & Hazard Control
- Wijs reagent is corrosive and lachrymatory; handle in a fume hood.
- Keep the reaction flask in the dark; light causes side reactions.
- Stopper flasks tightly; ICl is volatile.
Waste Disposal Protocol
Halogenated organic mixtures to halogenated solvent waste; thiosulphate titrates to aqueous waste.
Viva Voce Oral Defense Questions & Answers
Q: What does the iodine value tell you about an oil?
A: Its degree of unsaturation: higher IV means more C=C bonds per gram, distinguishing drying, semi-drying and non-drying oils.
Q: Why is the flask kept in the dark?
A: Light promotes substitution side reactions of ICl with the oil, which would inflate the apparent iodine absorption.
Q: Why add KI before titration?
A: KI converts unreacted ICl to I2 (ICl + KI -> KCl + I2), so the excess halogen can be measured by thiosulphate titration.