Spectrophotometric Determination of Iron as [Fe(phen)₃]²⁺
To determine Fe in a sample spectrophotometrically via [Fe(phen)₃]²⁺ at 510 nm.
Theoretical Foundations
Fe²⁺ + 3phen → [Fe(phen)₃]²⁺ (ε₅₁₀ ≈ 11,100). Fe(III) is first reduced with hydroxylamine. Beer's law is obeyed 0.5–5 ppm Fe. A calibration curve from standard Fe solutions gives the unknown concentration — a standard colorimetric method.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Cuvettes
- Volumetric flasks (100 mL)
- Pipettes
- Weighing balance
Chemical Reagents
- Iron standard (100 ppm Fe³⁺)
- Hydroxylamine hydrochloride (reductant)
- 1,10-phenanthroline (0.1%)
- Acetate buffer pH 4.5
- Unknown Fe sample
- Distilled water
Step-by-Step Practical Procedure
- Prepare standards: 0, 1, 2, 3, 4, 5 mL of 100 ppm Fe into 100 mL flasks.
- To each: add 2 mL hydroxylamine, 5 mL phenanthroline, 10 mL acetate buffer; dilute to mark.
- Wait 15 min for full colour development; read A₅₁₀.
- Plot calibration curve (linear through origin).
- Treat the unknown identically; read its concentration from the curve.
Observations & Expected Results
Orange-red colour proportional to Fe; calibration linear (r > 0.999); unknown determined to ±2%.
Worked Stoichiometric Calculations
Problem: Unknown A₅₁₀ = 0.444; calibration slope = 0.222 per ppm. Find Fe in ppm.
Working: c = 0.444/0.222 = 2.00 ppm Fe.
Safety Precautions & Hazard Control
- Hydroxylamine — gloves.
- Phenanthroline — gloves.
- Rinse cuvettes between readings.
Waste Disposal Protocol
Iron–phenanthroline waste to heavy-metal waste.
Viva Voce Oral Defense Questions & Answers
Q: Why hydroxylamine?
A: Reduces all Fe(III) to Fe(II), which is the species complexed.
Q: Why pH 4.5?
A: Optimal for complex formation; avoids Fe(OH)₃ precipitation.
Q: What is Beer's law?
A: A = εcl — absorbance proportional to concentration.