Zinc–EDTA Back-Titration: Determination of Al³⁺
To determine Al³⁺ by EDTA back-titration with Zn²⁺.
Theoretical Foundations
Al³⁺ reacts slowly with EDTA, so excess EDTA is added, boiled to complete complexation, and the excess is back-titrated with standard Zn²⁺ using Eriochrome Black T at pH 10. Al³⁺ = EDTA(total) − EDTA(excess). Back-titration is used when direct titration is slow or the end point is poor.
Required Apparatus & Chemicals
Laboratory Apparatus
- 50 mL burette
- 250 mL conical flask
- Pipettes
- Hot plate
- Measuring cylinder
- Dropper
Chemical Reagents
- Aluminium sample (alum solution)
- EDTA 0.01 M (standardised, excess)
- Zinc sulphate 0.01 M (standardised)
- Eriochrome Black T
- pH 10 buffer
- Distilled water
Step-by-Step Practical Procedure
- Pipette 25 mL Al³⁺ sample; add 25.00 mL (pipetted exactly) of 0.01 M EDTA.
- Boil 5 min to complete Al–EDTA complexation; cool.
- Add pH 10 buffer + Eriochrome Black T (blue — excess EDTA binds indicator? actually wine-red from Zn later; start blue).
- Titrate excess EDTA with 0.01 M Zn²⁺ to wine-red end point.
- Al³⁺ moles = (V_EDTA − V_Zn) × M.
Observations & Expected Results
Blue solution turns wine red at the end point (Zn–indicator complex); back-titre gives Al³⁺ accurately where direct titration fails.
Worked Stoichiometric Calculations
Problem: 25.00 mL of 0.01 M EDTA added; back-titre 8.40 mL of 0.01 M Zn²⁺. Find mg Al³⁺ (Al = 26.98).
Working: EDTA consumed by Al = (25.00 − 8.40) × 0.01 = 0.166 mmol = moles Al³⁺. Mass = 0.166 × 26.98 = 4.48 mg.
Safety Precautions & Hazard Control
- Boil to complete Al complexation — essential.
- pH 10 buffer — ventilation.
- Exact pipetting of EDTA aliquot.
Waste Disposal Protocol
Titration waste to aqueous heavy-metal waste.
Viva Voce Oral Defense Questions & Answers
Q: Why back-titration for Al³⁺?
A: Al–EDTA formation is slow; direct titration gives a fading end point.
Q: Why boil?
A: To drive the slow Al³⁺ + EDTA complexation to completion.
Q: What colour change is seen?
A: Blue (free indicator) → wine red (Zn–indicator) at the end point.