Lead Acetate Test: Sulfur-Containing Amino Acids
To detect sulfur-containing amino acids (cysteine, cystine) by alkaline hydrolysis to sulfide.
Theoretical Foundations
Boiling with strong alkali hydrolyses cysteine/cystine, releasing sulfide, which precipitates black lead sulfide with lead acetate. Methionine's thioether sulfur is not released, so it is negative.
Required Apparatus & Chemicals
Laboratory Apparatus
- Test tubes
- Boiling water bath
- Test tube holder
- Dropper
Chemical Reagents
- 40% NaOH
- Lead acetate solution (5%)
- 1% cysteine
- 1% methionine
- 1% albumin
Step-by-Step Practical Procedure
- Add 2 mL of cysteine, methionine and albumin to separate tubes.
- Add 2 mL of 40% NaOH to each and boil for 2 minutes.
- Cool slightly and add 5 drops of lead acetate solution.
- Observe black/brown precipitate of PbS in positive tubes.
- Note that methionine remains negative.
Observations & Expected Results
Cysteine and albumin give a black precipitate of lead sulfide; methionine stays clear, showing its sulfur is not labile.
Worked Stoichiometric Calculations
Problem: Cysteine (M = 121) releases one S atom per molecule as sulfide. What mass of PbS (M = 239) forms from 60.5 mg of cysteine?
Working: Moles cysteine = 60.5/121 = 0.50 mmol, giving 0.50 mmol PbS = 0.50 x 239 = 119.5 mg of PbS.
Safety Precautions & Hazard Control
- Lead salts are toxic; wear gloves and collect all waste.
- Boiling 40% NaOH is caustic; use a holder.
- Do not inhale vapours from the boiling tubes.
- Never pour lead waste down the drain.
Waste Disposal Protocol
All lead-containing waste collected as toxic heavy-metal waste.
Viva Voce Oral Defense Questions & Answers
Q: Why is methionine negative in this test?
A: Its sulfur is a thioether (-S-CH3), which resists alkaline hydrolysis, unlike the thiol/disulfide of cysteine/cystine.
Q: What is the black precipitate?
A: Lead sulfide, PbS, formed from sulfide released by alkaline hydrolysis.