Quantitative Seliwanoff: Fructose Estimation by Resorcinol

To estimate fructose colorimetrically using the Seliwanoff resorcinol reaction.

Level: UniversitySafety: Needs supervisionTime: 60 min
Aim of Experiment: To estimate fructose colorimetrically using the Seliwanoff resorcinol reaction.

Theoretical Foundations

Ketoses heated with resorcinol in HCl give a red complex (read at 480 nm) far faster than aldoses. With strict timing, absorbance is proportional to fructose concentration, allowing estimation in honey or fruit juice.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Spectrophotometer
  • Boiling water bath
  • Stopwatch
  • Test tubes
  • Pipettes

Chemical Reagents

  • Seliwanoff reagent (0.05% resorcinol in HCl)
  • Fructose standards (0-100 ug/mL)
  • Honey sample (diluted)
  • Distilled water

Step-by-Step Practical Procedure

  1. Pipette 2 mL of each fructose standard and diluted honey into tubes.
  2. Add 6 mL Seliwanoff reagent, mix, and immerse in the boiling bath.
  3. Heat for exactly 8 minutes, then cool rapidly in ice.
  4. Read absorbance at 480 nm within 10 minutes.
  5. Determine fructose from the standard curve and correct for dilution.

Observations & Expected Results

A red colour develops proportional to fructose; honey samples give strong readings consistent with their high fructose content.

Worked Stoichiometric Calculations

Problem: Calibration: A480 = 0.0092 x C (ug/mL). Honey diluted 1:5000 gives A = 0.644. Find the fructose % w/w in honey (density 1.4 g/mL).

Working: C = 0.644/0.0092 = 70 ug/mL measured. Original = 70 x 5000 = 350,000 ug/mL = 350 mg/mL. Per gram honey: 350/1.4 = 250 mg/g = 25% w/w fructose (typical honey ~38%; the demonstration dilution shows the calculation).

Safety Precautions & Hazard Control

  • Time the heating exactly; aldoses interfere with longer heating.
  • Cool rapidly to stop the reaction.
  • Hot HCl fumes; ventilate well.
  • Read promptly; the colour fades slowly.

Waste Disposal Protocol

Cool, neutralise acidic resorcinol waste with bicarbonate, then drain with water.

Viva Voce Oral Defense Questions & Answers

Q: Why must heating time be exact in the quantitative version?

A: Aldoses slowly give the same colour; fixed short heating keeps the signal specific to ketoses.

Q: At what wavelength is the complex read?

A: 480 nm, the absorption maximum of the red resorcinol-HMF complex.