Quantitative Catalase Assay by Permanganate Titration of Residual H2O2

To determine catalase activity by titrating the undecomposed H2O2 with standardised KMnO4.

Level: UniversitySafety: Needs supervisionTime: 90 min
Aim of Experiment: To determine catalase activity by titrating the undecomposed H2O2 with standardised KMnO4.

Theoretical Foundations

After a fixed reaction time, the reaction is stopped with H2SO4 and residual H2O2 titrated: 2 MnO4- + 5 H2O2 + 6 H+ -> 2 Mn2+ + 5 O2 + 8 H2O. The difference between blank and sample titres gives H2O2 decomposed, hence catalase units (micromol H2O2/min).

Required Apparatus & Chemicals

Laboratory Apparatus

  • Burette
  • Conical flasks
  • Water bath at 25 C
  • Pipettes
  • Stopwatch

Chemical Reagents

  • 0.05 M H2O2
  • Dilute catalase extract
  • 0.02 M KMnO4 (standardised)
  • 2 M H2SO4

Step-by-Step Practical Procedure

  1. Mix 10 mL H2O2 with 1 mL extract at 25 C; start timing.
  2. After exactly 5 min, stop with 5 mL H2SO4.
  3. Titrate with KMnO4 to the first permanent pink.
  4. Titrate a no-enzyme blank identically.
  5. Calculate H2O2 decomposed from the titre difference.
  6. Express as U/mL (micromol H2O2/min/mL).

Observations & Expected Results

Typical: sample titre markedly lower than blank; the pink end point is sharp in acid.

Worked Stoichiometric Calculations

Problem: Blank titre 25.0 mL, sample 15.0 mL of 0.02 M KMnO4; 1 mL extract, 5 min. Find U/mL.

Working: Difference 10.0 mL x 0.02 M = 0.20 mmol MnO4- = 0.50 mmol H2O2 decomposed in 5 min = 100 micromol/min = 100 U/mL (typical illustrative value).

Safety Precautions & Hazard Control

  • KMnO4 stains skin; H2SO4 is corrosive.
  • Stop all flasks at exactly the same time.
  • Standardise KMnO4 against oxalate.

Waste Disposal Protocol

Manganese waste to heavy-metal aqueous collection.

Viva Voce Oral Defense Questions & Answers

Q: Write the permanganate-peroxide titration reaction.

A: 2 MnO4- + 5 H2O2 + 6 H+ -> 2 Mn2+ + 5 O2 + 8 H2O.

Q: Why stop with sulphuric acid?

A: It denatures catalase instantly and provides the H+ the titration needs.

Q: Why is this better than gas collection?

A: It is quantitative and works at low substrate concentrations.