Serum Urea Estimation by the Urease-Berthelot Method
To estimate urea in serum using urease followed by the Berthelot ammonia colour reaction.
Theoretical Foundations
Urease hydrolyses urea to ammonia; the Berthelot reaction (phenol + hypochlorite, nitroprusside catalyst) converts ammonia to blue indophenol (580 nm). The method is specific and sensitive, used clinically for kidney-function testing (blood urea nitrogen).
Required Apparatus & Chemicals
Laboratory Apparatus
- Spectrophotometer (580 nm)
- Water bath at 37 C
- Pipettes
- Test tubes
Chemical Reagents
- Urea standard (40 mg/dL)
- Urease solution
- Phenol-nitroprusside reagent
- Alkaline hypochlorite reagent
- Serum sample
Step-by-Step Practical Procedure
- Incubate 0.02 mL serum/standard with urease 10 min at 37 C.
- Add phenol-nitroprusside then alkaline hypochlorite.
- Incubate 15 min for colour development.
- Read A580 vs blank.
- Calculate urea concentration.
- Convert to blood urea nitrogen if required (BUN = urea/2.14).
Observations & Expected Results
Typical: blue colour develops; normal serum urea ~15-40 mg/dL gives moderate absorbance.
Worked Stoichiometric Calculations
Problem: A580: standard (40 mg/dL) = 0.60, sample = 0.45. Find serum urea and BUN.
Working: Urea = 40 x 0.45/0.60 = 30 mg/dL. BUN = 30/2.14 = 14.0 mg/dL (typical normal values).
Safety Precautions & Hazard Control
- Phenol and hypochlorite are corrosive; wear gloves.
- Serum is a biohazard.
- Develop colour for the full time.
Waste Disposal Protocol
Phenolic waste to hazardous organic collection; serum to biohazard bags.
Viva Voce Oral Defense Questions & Answers
Q: What is the Berthelot reaction?
A: Ammonia + phenol + hypochlorite -> blue indophenol (nitroprusside-catalysed).
Q: Why measure BUN clinically?
A: It reflects kidney function and protein catabolism.
Q: Why use urease rather than direct ammonia assay?
A: Urease makes the assay specific for urea.