CHEMVEN

Protein Quantification by UV Absorbance at 280 nm

To estimate protein concentration from A280 using the extinction coefficient, without any colour reagent.

Level: UniversitySafety: SafeTime: 45 min
Aim of Experiment: To estimate protein concentration from A280 using the extinction coefficient, without any colour reagent.

Theoretical Foundations

Tryptophan and tyrosine absorb at 280 nm, so A280 is proportional to protein concentration (Beer-Lambert law). For BSA, A280 (1 mg/mL, 1 cm) ~ 0.66. Nucleic acid contamination inflates A280; the A280/A260 ratio flags it (pure protein ~1.8 for A260/A280 inverted: A280/A260 ~ 0.57 means check carefully).

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV spectrophotometer (280, 260 nm)
  • Quartz cuvettes
  • Micropipettes

Chemical Reagents

  • BSA solution of known concentration
  • Unknown protein sample
  • Buffer (blank)

Step-by-Step Practical Procedure

  1. Blank the spectrophotometer with buffer at 280 and 260 nm.
  2. Read A280 of BSA standard; calculate its specific absorbance.
  3. Read A280 and A260 of the unknown.
  4. Calculate concentration from A280.
  5. Check the A260/A280 ratio for nucleic acid contamination.
  6. Compare with a Bradford value if available.

Observations & Expected Results

Typical: A280 scales with concentration; a high A260/A280 (>1.0) warns of nucleic acid contamination inflating the estimate.

Worked Stoichiometric Calculations

Problem: Unknown A280 = 0.66 in 1 cm cuvette; BSA standard 1 mg/mL gives A280 = 0.66. Find concentration.

Working: C = 0.66/0.66 = 1.0 mg/mL (typical illustrative value).

Safety Precautions & Hazard Control

  • Use quartz cuvettes; glass absorbs UV.
  • Blank carefully; buffers can absorb at 280 nm.
  • Keep cuvettes scrupulously clean.

Waste Disposal Protocol

Protein solutions down the sink.

Viva Voce Oral Defense Questions & Answers

Q: Which residues absorb at 280 nm?

A: Tryptophan (mainly) and tyrosine; phenylalanine weakly.

Q: Why can A280 mislead?

A: Nucleic acids absorb strongly at 260-280 nm; also, Trp content varies per protein.

Q: What is the Beer-Lambert law?

A: A = epsilon x c x l.

← Back to Experiments Directory