CHEMVEN

Dissolved Organic Carbon Proxy: UV254 Absorbance (SUVA)

To estimate dissolved organic carbon (DOC) in water using UV absorbance at 254 nm (SUVA concept).

Level: UniversitySafety: SafeTime: 30 min
Aim of Experiment: To estimate dissolved organic carbon (DOC) in water using UV absorbance at 254 nm (SUVA concept).

Theoretical Foundations

Aromatic organic matter absorbs UV at 254 nm; UV254 absorbance correlates with DOC, and SUVA (UV254/DOC) indicates aromaticity and disinfection-by-product formation potential. Filtered samples are measured in quartz cuvettes. High-SUVA waters form more trihalomethanes on chlorination.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV spectrophotometer
  • Quartz cuvettes
  • Membrane filter 0.45 um
  • Filtration assembly
  • Wash bottle

Chemical Reagents

  • Potassium hydrogen phthalate (DOC standards, optional)
  • Distilled water
  • Dilute sulphuric acid (pH adjustment, if needed)

Step-by-Step Practical Procedure

  1. Filter the sample through a 0.45 um membrane.
  2. Zero the spectrophotometer with distilled water at 254 nm.
  3. Measure sample absorbance at 254 nm in a quartz cuvette.
  4. If DOC is known (TOC analyser), compute SUVA = UV254 x 100 / DOC.
  5. Use the site-specific UV254-DOC correlation to estimate DOC.
  6. Assess disinfection-by-product formation potential.

Observations & Expected Results

Humic-rich waters: UV254 0.1-0.5 cm-1; clear groundwater: <0.02 cm-1.

Worked Stoichiometric Calculations

Problem: A filtered sample has UV254 = 0.12 cm-1 and DOC = 4.0 mg/L. Calculate SUVA and interpret.

Working: SUVA = 0.12 x 100 / 4.0 = 3.0 L/mg-m. SUVA >2 indicates aromatic, humic-rich organics with significant trihalomethane formation potential on chlorination; enhanced coagulation is advisable.

Safety Precautions & Hazard Control

  • Filter before measurement; particles scatter UV.
  • Use quartz cuvettes.
  • Correct for nitrate interference if high (absorbs near 220 nm, minimal at 254).
  • Keep pH consistent; UV absorbance is pH-sensitive.

Waste Disposal Protocol

No hazardous waste; aqueous samples only.

Viva Voce Oral Defense Questions & Answers

Q: What is SUVA?

A: Specific UV absorbance = UV254 (per metre) divided by DOC (mg/L); a measure of aromatic carbon content.

Q: Why does high SUVA matter for chlorination?

A: Aromatic organics are the main precursors of trihalomethanes and haloacetic acids.

Q: Why filter at 0.45 um?

A: To measure dissolved organics only; particles would scatter light and inflate absorbance.

← Back to Experiments Directory