Total Suspended Solids: Gravimetric Determination
To determine total suspended solids (TSS) in wastewater by glass-fibre filtration and gravimetry.
Theoretical Foundations
A measured sample volume is filtered through a pre-weighed glass-fibre filter (0.7-1.2 um); the residue dried at 103-105 C is weighed. TSS (mg/L) = (weight gain in mg x 1000) / sample volume in mL. Discharge standards commonly limit TSS to 100 mg/L (inland surface waters); raw sewage is typically 150-400 mg/L.
Required Apparatus & Chemicals
Laboratory Apparatus
- Filtration assembly with vacuum pump
- Glass-fibre filter discs (pre-weighed)
- Drying oven (103 C)
- Desiccator
- Analytical balance
Chemical Reagents
- Distilled water (filter washing)
- Silica gel desiccant (for desiccator)
- Laboratory detergent (glassware cleaning)
Step-by-Step Practical Procedure
- Dry filter discs at 103-105 C, cool in a desiccator, and weigh (W1).
- Filter a well-mixed measured volume (100-500 mL) under vacuum.
- Wash the residue with distilled water.
- Dry the filter at 103-105 C for 1 hour.
- Cool in a desiccator and weigh (W2); repeat drying to constant weight.
- TSS = (W2 - W1) mg x 1000 / sample mL.
Observations & Expected Results
Visible residue on filter. Raw sewage: 150-400 mg/L; treated effluent: <30 mg/L typically.
Worked Stoichiometric Calculations
Problem: 250 mL of effluent gave a filter weight gain of 18.5 mg. Calculate TSS and check a 100 mg/L discharge limit.
Working: TSS = 18.5 x 1000 / 250 = 74 mg/L. Below the 100 mg/L discharge limit; compliant.
Safety Precautions & Hazard Control
- Shake the sample well; solids settle fast.
- Dry to constant weight.
- Handle filters with forceps, not fingers.
- Choose a sample volume giving 10-200 mg residue.
Waste Disposal Protocol
Dried filters with residue go to solid lab waste.
Viva Voce Oral Defense Questions & Answers
Q: Why glass-fibre and not paper filters?
A: Glass-fibre filters have a defined fine pore size, withstand 103 C drying, and give low, stable blank weights.
Q: Why dry at 103-105 C specifically?
A: Hot enough to remove water but not so hot as to volatilise organic matter or decompose bicarbonates significantly.
Q: How do TSS and turbidity differ?
A: TSS is a mass concentration (mg/L) by weighing; turbidity is an optical property (NTU) - related but not identical.