Free Carbon Dioxide in Water: Titrimetric Determination
To determine free (aggressive) carbon dioxide in water by titration with sodium carbonate.
Theoretical Foundations
Free CO2 is titrated with standard sodium carbonate to the phenolphthalein endpoint (pH 8.3), converting CO2 to bicarbonate. CO2 (mg/L) = V x N x 44000 / sample volume. Free CO2 makes water corrosive to concrete and metals; boiler feed and distribution systems monitor it.
Required Apparatus & Chemicals
Laboratory Apparatus
- Burette
- Conical flasks
- Pipettes
- Narrow-mouth sampling bottles
- White tile
Chemical Reagents
- Sodium carbonate (0.02 N)
- Phenolphthalein indicator
- Potassium hydrogen phthalate (standardisation)
Step-by-Step Practical Procedure
- Take 100 mL of sample with minimal aeration.
- Add phenolphthalein; if pink, free CO2 is absent.
- Titrate with 0.02 N Na2CO3 to the first permanent faint pink.
- Record the titre.
- Calculate CO2 = V x N x 44000 / 100 mg/L.
Observations & Expected Results
Groundwater: 5-50 mg/L free CO2; surface water in equilibrium with air: <5 mg/L.
Worked Stoichiometric Calculations
Problem: 100 mL of well water required 7.5 mL of 0.02 N Na2CO3. Calculate free CO2.
Working: CO2 = 7.5 x 0.02 x 44000 / 100 = 7.5 x 8.8 = 66 mg/L. High; the water is corrosive and would benefit from aeration (CO2 stripping).
Safety Precautions & Hazard Control
- Avoid aeration during sampling; CO2 escapes quickly.
- Titrate immediately after collection.
- Use a narrow-mouth bottle filled to the brim.
- Do not shake the sample.
Waste Disposal Protocol
Titrated solutions are harmless; drain disposal is fine.
Viva Voce Oral Defense Questions & Answers
Q: Why must the sample not be aerated?
A: Aeration strips dissolved CO2 into the air, giving falsely low results.
Q: Derive the factor 44000.
A: Equivalent weight of CO2 = 44.0/2 = 22 g/equiv (CO2 + CO3(2-) -> 2 HCO3-); mg/L = V x N x 22 x 1000 / V_sample.
Q: Why is free CO2 called aggressive?
A: It dissolves calcium carbonate (concrete, limestone) and corrodes iron pipes.