Arsenic in Groundwater: Silver Diethyldithiocarbamate Method
To determine trace arsenic in groundwater by arsine generation and SDDC colorimetry.
Theoretical Foundations
Arsenic is reduced to arsine gas (AsH3) by zinc in acid; the gas is scrubbed of H2S by lead acetate cotton and absorbed in silver diethyldithiocarbamate (SDDC) in pyridine, forming a red complex measured at 535 nm. WHO guideline: 0.01 mg/L; chronic exposure causes skin lesions and cancers. The method detects down to ~0.002 mg/L.
Required Apparatus & Chemicals
Laboratory Apparatus
- Arsine generator (Gutzeit-type bottle with scrubber tube)
- Spectrophotometer (535 nm)
- Cuvettes
- Fume hood
Chemical Reagents
- Zinc granules (arsenic-free)
- Hydrochloric acid (conc.)
- Lead acetate impregnated cotton
- SDDC reagent in pyridine
- Sodium arsenite stock standard
- Potassium iodide + stannous chloride (reductants)
Step-by-Step Practical Procedure
- Take 50 mL of acid-preserved sample in the generator bottle.
- Add KI and stannous chloride to reduce As(V) to As(III); wait 15 minutes.
- Place lead acetate cotton in the scrubber tube and SDDC-pyridine in the absorber.
- Add zinc granules and conc. HCl; connect the apparatus immediately.
- Allow 45 minutes for complete arsine evolution and absorption (red colour develops).
- Measure absorbance at 535 nm against a reagent blank.
- Read arsenic from a calibration curve run through the same generation step.
Observations & Expected Results
Yellow to red colour in the absorber. Arsenic-affected aquifers (Bengal delta type): 0.05-1.0 mg/L, far above the 0.01 mg/L WHO guideline.
Worked Stoichiometric Calculations
Problem: A 50 mL sample gives A = 0.30 at 535 nm; calibration A = 6.0 x C (mg/L As). Find arsenic and comment.
Working: C = 0.30/6.0 = 0.05 mg/L As. This is five times the WHO guideline of 0.01 mg/L; the well must not be used for drinking without arsenic removal treatment.
Safety Precautions & Hazard Control
- Arsine is extremely toxic; assemble quickly and work in a fume hood.
- The lead acetate scrubber is essential to remove H2S, which also colours SDDC.
- Use arsenic-free zinc and acids; run blanks.
- Pyridine is toxic and foul-smelling; handle in the fume hood.
Waste Disposal Protocol
SDDC-pyridine absorber solutions and generator residues go to the hazardous waste container.
Viva Voce Oral Defense Questions & Answers
Q: What is the function of the lead acetate cotton?
A: It scrubs hydrogen sulphide from the gas stream; H2S would also react with SDDC and give false high arsenic.
Q: Why are KI and stannous chloride added first?
A: They pre-reduce As(V) to As(III), which is the form that generates arsine with zinc and acid.
Q: Name a chronic health effect of arsenic in drinking water.
A: Hyperkeratosis/skin lesions, and long-term cancers of skin, bladder and lung.