CHEMVEN

Hexavalent Chromium: Diphenylcarbazide Method

To determine hexavalent chromium in industrial effluent by the diphenylcarbazide method.

Level: UniversitySafety: Advanced lab onlyTime: 60 min
Aim of Experiment: To determine hexavalent chromium in industrial effluent by the diphenylcarbazide method.

Theoretical Foundations

Cr(VI) reacts with 1,5-diphenylcarbazide in acid to form a red-violet complex measured at 540 nm; the method is specific for the toxic hexavalent form. WHO drinking-water guideline for total chromium is 0.05 mg/L; Cr(VI) is carcinogenic. Tannery and electroplating effluents are the main sources, with discharge limits commonly 0.1 mg/L Cr(VI).

Required Apparatus & Chemicals

Laboratory Apparatus

  • Spectrophotometer (540 nm)
  • Cuvettes
  • Volumetric flasks
  • pH meter

Chemical Reagents

  • 1,5-diphenylcarbazide in acetone
  • Sulphuric acid (dilute)
  • Potassium dichromate stock (Cr(VI) standard)

Step-by-Step Practical Procedure

  1. Filter the sample if turbid.
  2. Adjust an aliquot to pH 1-2 with dilute sulphuric acid.
  3. Add diphenylcarbazide reagent and mix.
  4. Wait 5-10 minutes for full colour development.
  5. Measure absorbance at 540 nm against a reagent blank.
  6. Read Cr(VI) from a calibration curve prepared with dichromate standards.

Observations & Expected Results

Red-violet colour proportional to Cr(VI). Electroplating effluents: 10-100 mg/L before treatment.

Worked Stoichiometric Calculations

Problem: A sample gives A = 0.54 at 540 nm; calibration A = 2.70 x C (mg/L Cr(VI)). Find Cr(VI) and check a 0.1 mg/L discharge limit.

Working: C = 0.54/2.70 = 0.20 mg/L Cr(VI). This is twice the 0.1 mg/L discharge limit; reduction to Cr(III) and precipitation treatment is required.

Safety Precautions & Hazard Control

  • Cr(VI) standards are carcinogenic; wear gloves and avoid dust.
  • Develop colour at the correct pH (1-2).
  • Read within 30 minutes; the complex fades slowly.
  • Reduce Cr(VI) waste with ferrous sulphate before disposal.

Waste Disposal Protocol

Reduce all Cr(VI) waste to Cr(III) with ferrous sulphate in acid, neutralise, and collect the sludge as hazardous waste.

Viva Voce Oral Defense Questions & Answers

Q: Why is the method specific for Cr(VI)?

A: Only hexavalent chromium oxidises diphenylcarbazide to form the coloured complex; Cr(III) does not react.

Q: How is Cr(VI) removed from effluents?

A: By reduction to Cr(III) with SO2 or FeSO4 at low pH, then precipitation as Cr(OH)3 with lime at pH 8-9.

Q: Why is Cr(VI) more hazardous than Cr(III)?

A: Cr(VI) is a strong oxidant, readily crosses cell membranes, and is carcinogenic; Cr(III) is poorly absorbed and an essential trace nutrient.

← Back to Experiments Directory