Coliform Bacteria: Most Probable Number (MPN) Test
To estimate coliform bacteria in drinking water by the multiple-tube (MPN) fermentation technique.
Theoretical Foundations
Coliforms are indicator organisms for faecal contamination. In the MPN test, measured sample volumes are inoculated into lactose broth with Durham tubes; gas production within 48 h at 37 C indicates coliforms. The pattern of positive tubes is matched to McCrady probability tables to give the MPN per 100 mL. WHO guideline: zero E. coli/thermotolerant coliforms per 100 mL of drinking water.
Required Apparatus & Chemicals
Laboratory Apparatus
- Test tubes with Durham tubes
- Autoclave
- Incubator at 37 C
- Pipettes (sterile)
- Bunsen burner
Chemical Reagents
- MacConkey/lactose broth (single and double strength)
- Distilled water (sterile dilution blanks)
- Brilliant green bile broth (confirmation)
Step-by-Step Practical Procedure
- Sterilise all tubes, broth and pipettes by autoclaving.
- Inoculate 5 tubes of double-strength broth with 10 mL sample each, 5 tubes of single-strength with 1 mL, and 5 with 0.1 mL.
- Incubate at 37 C for 48 hours.
- Record tubes showing gas in Durham tubes and acid (colour change) as positive.
- Look up the 5-5-5 positive pattern in McCrady MPN tables per 100 mL.
- Confirm faecal coliforms by subculture at 44.5 C if required.
Observations & Expected Results
Gas bubbles in Durham tubes; broth turns yellow. Safe drinking water: 0 coliforms/100 mL; polluted wells: hundreds per 100 mL.
Worked Stoichiometric Calculations
Problem: A 5-5-5 tube set gives positives 4-2-0. The MPN table value is 22 per 100 mL. Is this water safe per WHO?
Working: MPN = 22 coliforms per 100 mL. WHO requires zero E. coli per 100 mL in drinking water, so this water is unsafe and must be disinfected (boiling/chlorination) before drinking.
Safety Precautions & Hazard Control
- Maintain strict aseptic technique to avoid false positives.
- Sterilise all contaminated material by autoclaving before disposal.
- Incubate at the correct temperature; 44.5 C confirms faecal coliforms.
- Use samples within 6 hours of collection, kept cool.
Waste Disposal Protocol
Autoclave all positive tubes and cultures at 121 C before disposal as biohazard waste.
Viva Voce Oral Defense Questions & Answers
Q: Why are coliforms used as indicators rather than testing for pathogens directly?
A: Coliforms are abundant in faeces, easy to culture, and their presence reliably signals possible pathogen contamination.
Q: What is the significance of gas in the Durham tube?
A: Gas (CO2 + H2) from lactose fermentation is the definitive presumptive sign of coliforms.
Q: What does the MPN value actually represent?
A: A statistical estimate, not an exact count, of organism density based on the pattern of positive tubes.