CHEMVEN

Fluoride in Water: SPADNS Colorimetric Method

To determine fluoride in drinking water by the SPADNS spectrophotometric method.

Level: UniversitySafety: Needs supervisionTime: 60 min
Aim of Experiment: To determine fluoride in drinking water by the SPADNS spectrophotometric method.

Theoretical Foundations

Fluoride reacts with a red zirconium-SPADNS complex, bleaching it in proportion to fluoride concentration (the dye-lake is destroyed). Absorbance is measured at 570 nm and compared with a calibration curve. The method covers 0-1.4 mg/L. WHO guideline: 1.0 mg/L (1.5 mg/L maximum); excess fluoride causes dental and skeletal fluorosis, while too little increases dental caries.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Spectrophotometer (570 nm)
  • Cuvettes
  • Volumetric flasks 50 mL
  • Pipettes
  • Polythene bottles for samples

Chemical Reagents

  • SPADNS reagent
  • Zirconyl acid reagent
  • Sodium fluoride stock (100 mg/L F-)
  • Distilled water (fluoride-free)

Step-by-Step Practical Procedure

  1. Prepare fluoride standards 0, 0.2, 0.4, 0.8, 1.0, 1.4 mg/L in 50 mL flasks.
  2. Add SPADNS and zirconyl reagents to standards and samples; mix and wait for colour development.
  3. Set the spectrophotometer to 570 nm and zero with the reagent blank.
  4. Measure absorbance of standards and plot the calibration curve (absorbance falls as fluoride rises).
  5. Measure the sample absorbance and read its fluoride from the curve.
  6. Compare with the WHO guideline of 1.0 mg/L.

Observations & Expected Results

Red colour bleaches progressively with fluoride. Fluoride-belt groundwater in many regions: 2-10 mg/L (fluorosis risk).

Worked Stoichiometric Calculations

Problem: A sample gives absorbance 0.42 at 570 nm. The calibration line is A = 0.80 - 0.38 x C (C in mg/L). Find fluoride and comment.

Working: 0.42 = 0.80 - 0.38C, so 0.38C = 0.38 and C = 1.0 mg/L. This equals the WHO guideline value of 1.0 mg/L - acceptable, but any higher would need defluoridation.

Safety Precautions & Hazard Control

  • Use polythene, not glass, for storing fluoride samples (glass adsorbs fluoride).
  • Control temperature; the bleaching reaction is temperature-sensitive.
  • Remove aluminium and phosphate interferences by distillation if present in high amounts.
  • Prepare the combined reagent fresh.

Waste Disposal Protocol

Zirconium-containing waste goes to the inorganic chemical waste container.

Viva Voce Oral Defense Questions & Answers

Q: Why does absorbance decrease as fluoride increases in this method?

A: Fluoride destroys the coloured zirconium-dye complex, so more fluoride means less colour and lower absorbance.

Q: What are the health effects of excess fluoride?

A: Dental fluorosis (mottled teeth) above ~1.5 mg/L and skeletal fluorosis (joint stiffness) at much higher chronic intake.

Q: Why store fluoride samples in polythene?

A: Glass surfaces adsorb fluoride ions, giving low results; polythene avoids this loss.