Potentiometric Argentometry: Chloride with Silver Electrode
To determine chloride by potentiometric titration with AgNO3, independent of indicators.
Theoretical Foundations
A silver wire electrode responds to [Ag+] via Nernst: E = E0 + 0.05916 log[Ag+]. As AgCl precipitates, [Ag+] is fixed by Ksp; at equivalence it jumps ~200 mV. The first-derivative peak gives Veq. Works in coloured/turbid samples where Mohr/Fajans fail.
Required Apparatus & Chemicals
Laboratory Apparatus
- mV meter
- Ag wire + reference electrodes
- 50 mL burette
- 250 mL beaker
- Magnetic stirrer
- 50 mL pipette
Chemical Reagents
- Chloride sample (unknown)
- Standardised AgNO3 (0.01004 M)
- Distilled water
Step-by-Step Practical Procedure
- Pipette 50.00 mL sample into a beaker.
- Insert electrodes; stir.
- Add AgNO3 in 0.50 mL steps, recording mV.
- Use 0.10 mL steps across the jump.
- Plot E and dE/dV vs V; read Veq.
- Calculate mg/L Cl.
Observations & Expected Results
Potential jump of about 200 mV near 14.20 mL; derivative peak sharp.
Worked Stoichiometric Calculations
Problem: 50.00 mL sample; derivative peak at 14.20 mL of 0.01004 M AgNO3. Find mg/L Cl (M = 35.45).
Working: Moles Ag+ = 0.01004 x 0.01420 = 0.00014257 mol = moles Cl-. Mass = 0.00014257 x 35.45 = 0.0050540 g per 50 mL = 101 mg/L.
Safety Precautions & Hazard Control
- Polish the Ag electrode if the jump is rounded.
- Stir at constant speed.
- Shield from light.
Waste Disposal Protocol
Silver waste to silver recovery.
Viva Voce Oral Defense Questions & Answers
Q: Why does the potential jump at equivalence?
A: [Ag+] is pinned low by excess Cl- before equivalence and rises sharply once Ag+ is in excess, per the Nernst equation.
Q: Name an advantage over Mohr's method.
A: No indicator or colour judgment is needed, so turbid, coloured or very dilute samples can be titrated accurately.