Iodate in Iodised Salt: Iodometric Assay

To determine potassium iodate (mg/kg) in iodised table salt.

Level: UniversitySafety: SafeTime: 60 min
Aim of Experiment: To determine potassium iodate (mg/kg) in iodised table salt.

Theoretical Foundations

IO3(-) + 5I- + 6H+ -> 3I2 + 3H2O; I2 titrated with very dilute thiosulfate. Salt is iodised with KIO3 (stable) rather than KI (oxidises). The method detects mg/kg levels using 0.005 M thiosulfate.

Required Apparatus & Chemicals

Laboratory Apparatus

  • 50 mL burette
  • 100 mL volumetric flask
  • 25 mL pipette
  • 250 mL iodine flask
  • Analytical balance

Chemical Reagents

  • Iodised salt
  • KI crystals
  • Dilute HCl
  • Standardised Na2S2O3 (0.005000 M)
  • Starch indicator

Step-by-Step Practical Procedure

  1. Weigh 10.00 g salt; dissolve and make up to 100.0 mL.
  2. Pipette 25.00 mL into an iodine flask.
  3. Add 1 g KI and 5 mL dilute HCl; keep dark 5 min.
  4. Titrate with 0.005000 M thiosulfate, starch near the end.
  5. Calculate mg/kg as KIO3 (M = 214.00).

Observations & Expected Results

Faint brown I2; blue to colourless endpoint. Typical titre about 0.90 mL.

Worked Stoichiometric Calculations

Problem: 10.00 g salt to 100.0 mL; 25.00 mL aliquot needed 0.90 mL of 0.005000 M thiosulfate. Find mg/kg as KIO3.

Working: Moles thio = 0.005000 x 0.00090 = 0.0000045000 mol. Moles I2 = half = 0.0000022500 mol. Moles IO3(-) = /3 = 0.00000075000 mol in aliquot. In 100 mL: 0.0000030000 mol. Mass KIO3 = 0.0000030000 x 214.00 = 0.00064200 g per 10 g = 64.2 mg/kg.

Safety Precautions & Hazard Control

  • Use very dilute thiosulfate; the iodine level is tiny.
  • A reagent blank should be negligible; check it.
  • Dissolve the salt completely.

Waste Disposal Protocol

Dilute iodide solutions may be flushed with water.

Viva Voce Oral Defense Questions & Answers

Q: Why is KIO3 used for iodisation rather than KI?

A: KI is slowly oxidised by air to I2, which sublimes away; KIO3 is stable indefinitely in dry salt.

Q: How many thiosulfate moles per iodate?

A: Six: IO3(-) -> 3 I2, each I2 takes 2 S2O3(2-).

← Back to Experiments Directory