CHEMVEN

Gran Plot: Linearised Potentiometric Endpoint

To locate a potentiometric equivalence point by Gran linearisation using data before the endpoint only.

Level: UniversitySafety: SafeTime: 75 min
Aim of Experiment: To locate a potentiometric equivalence point by Gran linearisation using data before the endpoint only.

Theoretical Foundations

Before equivalence in a strong acid-strong base titration, [H+] = (Va - Vb)Cb/(V0 + Vb). A Gran plot of (V0 + V) x 10^(-pH) vs V is linear and extrapolates to zero at Veq. It uses only pre-equivalence data, avoiding the noisy jump region, and is more precise than derivative methods.

Required Apparatus & Chemicals

Laboratory Apparatus

  • pH meter
  • 50 mL burette
  • 250 mL beaker
  • Magnetic stirrer
  • 25 mL pipette
  • Graph paper or spreadsheet

Chemical Reagents

  • HCl (approx 0.1 M)
  • Standardised NaOH (approx 0.1 M)
  • pH buffers
  • Distilled water

Step-by-Step Practical Procedure

  1. Calibrate the pH meter.
  2. Pipette 25.00 mL HCl into a beaker; add water to about 100 mL.
  3. Add NaOH in 2.00 mL steps, recording pH, stopping before the jump.
  4. Compute G = (V0 + V) x 10^(-pH) for each point.
  5. Plot G vs V; extrapolate the straight line to G = 0.
  6. Read Veq from the x-intercept and calculate M(HCl).

Observations & Expected Results

Gran plot is strikingly linear; x-intercept at about 24.55 mL.

Worked Stoichiometric Calculations

Problem: Gran x-intercept 24.55 mL of 0.1000 M NaOH for 25.00 mL HCl. Find M(HCl).

Working: Moles NaOH at Veq = 0.1000 x 0.02455 = 0.002455 mol = moles HCl. M(HCl) = 0.002455/0.02500 = 0.09820 mol/L.

Safety Precautions & Hazard Control

  • Use only points well before the jump; near-equivalence points curve.
  • Accurate pH values matter because 10^(-pH) amplifies error.
  • Keep ionic strength roughly constant.

Waste Disposal Protocol

Neutral solutions may be flushed with water.

Viva Voce Oral Defense Questions & Answers

Q: Why is the Gran plot linear before equivalence?

A: Because (V0+V)[H+] is proportional to unreacted acid (Va - V), a linear function of added volume V.

Q: What is the main advantage over the derivative method?

A: It avoids the steep jump region where readings are hardest to take accurately, and extrapolation averages random error.

← Back to Experiments Directory