CHEMVEN

Potentiometric Titration: Strong Acid versus Strong Base

To locate the equivalence point of an HCl-NaOH titration from pH-volume data using first and second derivatives.

Level: UniversitySafety: SafeTime: 90 min
Aim of Experiment: To locate the equivalence point of an HCl-NaOH titration from pH-volume data using first and second derivatives.

Theoretical Foundations

A glass electrode follows the Nernst equation: E = E0 - 0.05916 pH at 25 C. Plotting pH vs volume gives a sigmoid; the equivalence point is the inflection, found as the maximum of delta-pH/delta-V or the zero of the second derivative. No indicator is needed, so coloured or turbid samples can be titrated.

Required Apparatus & Chemicals

Laboratory Apparatus

  • pH meter with glass + calomel electrodes
  • 50 mL burette
  • 250 mL beaker
  • Magnetic stirrer
  • 25 mL pipette

Chemical Reagents

  • HCl solution (approx 0.1 M)
  • Standardised NaOH (approx 0.1 M)
  • pH 4.00 and 7.00 buffer sachets
  • Distilled water

Step-by-Step Practical Procedure

  1. Calibrate the pH meter with pH 7.00 then 4.00 buffers.
  2. Pipette 25.00 mL HCl into a beaker; add water to cover electrodes.
  3. Add NaOH in 1.00 mL steps, recording pH after stabilisation.
  4. Near the jump (pH 4 to 10), add 0.20 mL steps.
  5. Continue 5 mL past the jump.
  6. Plot pH vs V, compute first/second derivatives, find equivalence volume.

Observations & Expected Results

pH rises slowly, jumps about 6 units within 0.4 mL, then levels. Second derivative crosses zero at about 24.55 mL.

Worked Stoichiometric Calculations

Problem: 25.00 mL HCl; second-derivative zero at 24.55 mL of 0.1000 M NaOH. Find M(HCl).

Working: 1:1 stoichiometry. Moles NaOH = 0.1000 x 0.02455 = 0.002455 mol = moles HCl. M(HCl) = 0.002455/0.02500 = 0.09820 mol/L.

Safety Precautions & Hazard Control

  • Calibrate the meter before use; rinse electrodes between readings.
  • Stir uniformly and wait for a stable reading.
  • Keep temperature constant; pH is temperature-dependent.

Waste Disposal Protocol

Neutral salt solutions may be flushed with water.

Viva Voce Oral Defense Questions & Answers

Q: Why is the second derivative zero at equivalence?

A: Equivalence is the inflection point of the sigmoid: maximum slope (first derivative peak), hence the second derivative changes sign there.

Q: Name one advantage of potentiometry over indicators.

A: It works in coloured, turbid or fluorescent solutions where a colour change cannot be seen.

← Back to Experiments Directory