Spectrophotometric Determination of Iron(II/III) using 1,10-Phenanthroline
To determine trace concentrations of iron in an unknown water/ore sample by measuring the absorbance of the orange-red tris(1,10-phenanthroline)iron(II) complex, [Fe(phen)3]2+, at 510 nm and applying the Beer-Lambert law.
Theoretical Foundations
Iron in solution is reduced to iron(II) using hydroxylamine hydrochloride: 2Fe3+ + 2NH2OH·HCl -> 2Fe2+ + N2 + 2H2O + 4H+ + 2Cl-. The Fe2+ ions react stoichiometrically with three molecules of 1,10-phenanthroline (orthophenanthroline) to form the intense orange-red octahedral tris(1,10-phenanthroline)iron(II) chelate ("ferroin"): Fe2+ + 3(phen) -> [Fe(phen)3]2+. The complex exhibits intense visible light absorption with a sharp maximum (lambda_max) at 510 nm and a high molar absorptivity (epsilon ≈ 11,100 L·mol^-1·cm^-1). The complex obeys the Beer-Lambert law, A = epsilon·b·c, over the concentration range 0.2 to 5.0 mg/L Fe across a pH range of 3.5 to 4.5 maintained with sodium acetate buffer.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Visible spectrophotometer (digital, 320-1000 nm)
- Matched 1.00 cm quartz or optical glass cuvettes
- 100 mL volumetric flasks (six)
- 10 mL volumetric pipettes
- Graduated cylinders (10 mL, 25 mL)
- Analytical balance (±0.0001 g)
Chemical Reagents
- Standard iron stock solution (100 mg/L Fe from pure ferrous ammonium sulphate)
- Hydroxylamine hydrochloride solution (10% w/v, NH2OH·HCl)
- 1,10-Phenanthroline monohydrate solution (0.25% w/v in water)
- Sodium acetate buffer solution (2.0 M, pH ~4.5)
- Unknown iron sample solution
- Deionised water
Step-by-Step Practical Procedure
- Prepare a 10.0 mg/L working iron standard by diluting 10.00 mL of the 100 mg/L stock to 100.0 mL with deionised water.
- Into a series of six 100 mL volumetric flasks, pipette 0.00 (blank), 2.00, 4.00, 6.00, 8.00, and 10.00 mL of the working iron standard (corresponding to 0.0, 0.2, 0.4, 0.6, 0.8, and 1.0 mg/L Fe).
- Pipette 10.00 mL of the unknown iron sample into a seventh 100 mL volumetric flask.
- To each flask, add in order: 2.0 mL of 10% hydroxylamine hydrochloride solution, 5.0 mL of sodium acetate buffer, and 5.0 mL of 0.25% 1,10-phenanthroline solution.
- Dilute each flask to the 100.0 mL mark with deionised water, stopper, invert 5 times, and allow 15 minutes for full color development.
- Switch on the spectrophotometer, set the wavelength to 510 nm, and zero the instrument using the reagent blank in a 1.00 cm cuvette.
- Measure the absorbance (A) of each standard solution in order of increasing concentration, and then measure the unknown sample.
- Construct a calibration curve of Absorbance vs Concentration (mg/L Fe); perform linear regression to determine slope and intercept.
- Interpolate the concentration of iron in the unknown flask, and account for dilution to calculate the iron concentration of the original sample.
Observations & Expected Results
Solutions develop an intense, sparkling orange-red color within 5 minutes. Absorbance increases linearly with iron concentration: 0.2 mg/L gives A ≈ 0.040, up to 1.0 mg/L giving A ≈ 0.200 (at 100 mL volume). The unknown sample absorbance falls precisely on the linear calibration line (R^2 > 0.9995).
Worked Stoichiometric Calculations
Problem: Calibration data at 510 nm (b = 1.0 cm): 0.20 mg/L -> A = 0.040; 0.40 mg/L -> A = 0.080; 0.60 mg/L -> A = 0.121; 0.80 mg/L -> A = 0.160; 1.00 mg/L -> A = 0.201. A 10.0 mL aliquot of unknown water diluted to 100.0 mL gave A = 0.145. Calculate the concentration of Fe in the original unknown water.
Working: Slope m = Delta A / Delta C = (0.201 - 0.040) / (1.00 - 0.20) = 0.161 / 0.80 = 0.201 L/mg. Intercept is 0.000. Concentration in measured flask C_flask = A / m = 0.145 / 0.201 = 0.721 mg/L Fe. Dilution factor = 100.0 mL / 10.0 mL = 10. Original Fe concentration = 0.721 mg/L × 10 = 7.21 mg/L Fe.
Safety Precautions & Hazard Control
- Always add reagents in the exact order: hydroxylamine first (to reduce all Fe to Fe2+), then buffer (to fix pH at 4.5), and then phenanthroline.
- Handle cuvettes only by their frosted sides; clean optical faces with lint-free lens tissue before inserting into the cell holder.
- Rinse cuvette twice with the test solution before taking absorbance readings.
Waste Disposal Protocol
Used phenanthroline-iron solutions may be disposed of in the general aqueous chemical waste collection container.
Viva Voce Oral Defense Questions & Answers
Q: Why is hydroxylamine hydrochloride added to the sample?
A: Iron in water samples often exists in both Fe2+ and Fe3+ oxidation states. 1,10-phenanthroline reacts selectively with Fe2+ to form the intense orange complex; hydroxylamine acts as a mild reducing agent that rapidly and quantitatively reduces all Fe3+ to Fe2+.
Q: Why is sodium acetate buffer added and what is the optimal pH range?
A: The optimal pH for [Fe(phen)3]2+ complex formation is 3.5 to 5.5. At pH < 3, phenanthroline becomes protonated (phenH+) and cannot coordinate effectively. At pH > 6, iron(II) precipitates as hydroxide. Sodium acetate maintains the solution at pH ~4.5.
Q: State the Beer-Lambert law and define each parameter.
A: A = epsilon · b · c, where A is absorbance (dimensionless), epsilon is molar absorptivity (L·mol^-1·cm^-1), b is the optical path length of the cuvette (cm), and c is the molar concentration of the absorbing solute (mol/L).