Fluorescence Quenching: Stern-Volmer Study of Quinine by Halide

To determine the Stern-Volmer quenching constant for quinine fluorescence quenched by iodide.

Level: UniversitySafety: SafeTime: 90 min
Aim of Experiment: To determine the Stern-Volmer quenching constant for quinine fluorescence quenched by iodide.

Theoretical Foundations

Collisional quenching follows the Stern-Volmer equation F0/F = 1 + Ksv[Q], where F0 and F are intensities without and with quencher. A plot of F0/F versus [I⁻] is linear with slope Ksv, the quenching constant, reflecting quenching efficiency.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Spectrofluorometer
  • Quartz cuvettes
  • Volumetric flasks
  • Pipettes

Chemical Reagents

  • Quinine sulfate (1 mg L⁻¹ in 0.05 M H2SO4)
  • Potassium iodide (1.0 M quencher)
  • Sulfuric acid (0.05 M)

Step-by-Step Practical Procedure

  1. Prepare quinine solutions with [I⁻] = 0-0.20 M (constant quinine).
  2. Set ex 350 nm / em 450 nm.
  3. Read F0 (no quencher) and F for each [Q].
  4. Plot F0/F versus [I⁻].
  5. Find Ksv from the slope.
  6. Distinguish collisional from static quenching by discussion.

Observations & Expected Results

Blue fluorescence dims progressively with iodide; F0/F versus [I⁻] is linear.

Worked Stoichiometric Calculations

Problem: F0 = 100; at [I⁻] = 0.10 M, F = 50. Find Ksv.

Working: F0/F = 2.0 = 1 + Ksv(0.10) → Ksv = 1.0/0.10 = 10 M⁻¹.

Safety Precautions & Hazard Control

  • Keep quinine concentration identical in all solutions.
  • Control temperature; collisional quenching is temperature-dependent.
  • Correct for dilution by the quencher additions.
  • Use fresh KI; I2 formation would add static quenching.

Waste Disposal Protocol

Dilute iodide/quinine waste to the aqueous drain.

Viva Voce Oral Defense Questions & Answers

Q: What does Ksv measure?

A: The efficiency of collisional quenching: larger Ksv means fewer collisions needed to quench, i.e. stronger quenching.

Q: How would temperature affect Ksv for collisional quenching?

A: It increases Ksv, because faster diffusion raises the collision rate.

Q: How could you detect static quenching?

A: Static quenching forms a non-fluorescent ground-state complex; absorption spectra change and lifetime measurements show no effect, unlike collisional quenching.