Fluorimetry: Quinine Sulfate Calibration (Ex 350 / Em 450 nm)

To build a fluorescence calibration for quinine sulfate and determine an unknown.

Level: UniversitySafety: SafeTime: 60 min
Aim of Experiment: To build a fluorescence calibration for quinine sulfate and determine an unknown.

Theoretical Foundations

Quinine in dilute H2SO4 fluoresces strongly (excitation 350 nm, emission 450 nm). At low concentrations fluorescence intensity is proportional to concentration (F = k·c), giving calibrations ~1000× more sensitive than absorbance. Inner-filter effects curve the plot at high concentrations.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Spectrofluorometer
  • Quartz cuvettes
  • Volumetric flasks
  • Pipettes

Chemical Reagents

  • Quinine sulfate standard (100 mg L⁻¹ in 0.05 M H2SO4)
  • Sulfuric acid (0.05 M)
  • Unknown quinine sample

Step-by-Step Practical Procedure

  1. Prepare quinine standards 0-1 mg L⁻¹ in 0.05 M H2SO4.
  2. Set excitation 350 nm, emission 450 nm.
  3. Zero with the acid blank.
  4. Read fluorescence intensities.
  5. Plot F versus concentration.
  6. Read the unknown from the linear region.

Observations & Expected Results

Blue fluorescence visible under UV; F linear to ~1 mg L⁻¹; typical slope ≈ 900 intensity units·L mg⁻¹.

Worked Stoichiometric Calculations

Problem: Slope 900 units·L mg⁻¹. An unknown gives F = 450. Find quinine in mg L⁻¹.

Working: c = 450/900 = 0.50 mg L⁻¹.

Safety Precautions & Hazard Control

  • Work in the linear low-concentration region.
  • Quinine photodegrades; minimize UV exposure.
  • Use quartz cuvettes.
  • Keep acid concentration constant; fluorescence is pH-sensitive.

Waste Disposal Protocol

Dilute acidic quinine waste to the aqueous drain.

Viva Voce Oral Defense Questions & Answers

Q: Why is fluorimetry more sensitive than spectrophotometry?

A: It measures light against a dark background (zero blank signal), while absorbance measures a small difference between two bright beams.

Q: What is the inner-filter effect?

A: At high concentration the sample absorbs its own excitation/emission light, curving the calibration downward.

Q: Why dilute sulfuric acid medium?

A: Quinine fluoresces strongly only as the protonated dication in acid.