Graphite Furnace AAS: Trace Lead in Water at 283.3 nm
To determine trace lead in water by graphite furnace AAS at 283.3 nm.
Theoretical Foundations
Electrothermal (graphite furnace) AAS atomizes the whole microlitre sample in a confined tube, giving ~100× better detection limits than flame AAS (~0.1 µg L⁻¹ for Pb). Programmed drying, ashing and atomization steps separate matrix removal from atomization; the 283.3 nm Pb line is monitored with Zeeman background correction.
Required Apparatus & Chemicals
Laboratory Apparatus
- GFAAS spectrometer
- Pb hollow-cathode lamp
- Autosampler
- Micropipettes
- Argon supply
Chemical Reagents
- Pb standard (100 µg L⁻¹)
- Nitric acid (ultrapure)
- Matrix modifier (NH4H2PO4/Mg(NO3)2)
- Water sample
Step-by-Step Practical Procedure
- Acidify sample with ultrapure HNO3.
- Prepare Pb standards 0-100 µg L⁻¹.
- Program: dry 110 °C, ash 600 °C, atomize 1800 °C.
- Inject 20 µL with matrix modifier.
- Record peak-area absorbances.
- Determine Pb; compare with the 10 µg L⁻¹ guideline.
Observations & Expected Results
Sharp transient peaks; peak area linear with Pb; typical slope ≈ 3.6 A·L mg⁻¹ (peak area).
Worked Stoichiometric Calculations
Problem: Slope 3.6 A·L mg⁻¹ (peak area). A sample gives A = 0.180. Find Pb in mg L⁻¹ and µg L⁻¹.
Working: c = 0.180/3.6 = 0.050 mg L⁻¹ = 50 µg L⁻¹ Pb (above the 10 µg L⁻¹ guideline).
Safety Precautions & Hazard Control
- Lead is toxic; micropipette carefully.
- Ultrapure reagents; contamination dominates at µg L⁻¹.
- Never exceed the tube's temperature program.
- Argon must flow to protect the graphite tube.
Waste Disposal Protocol
Trace Pb waste to the lead waste bottle.
Viva Voce Oral Defense Questions & Answers
Q: Why is GFAAS ~100× more sensitive than flame AAS?
A: The entire sample is atomized in the light path for seconds, versus milliseconds of dilute flame gases, hugely increasing atom density.
Q: What is the matrix modifier for?
A: It stabilizes Pb during ashing so matrix can be burnt off without losing analyte.
Q: Why peak area rather than height?
A: Area integrates the whole atomization pulse and is less affected by atomization kinetics.