CHEMVEN

Graphite Furnace AAS: Trace Lead in Water at 283.3 nm

To determine trace lead in water by graphite furnace AAS at 283.3 nm.

Level: UniversitySafety: Advanced lab onlyTime: 2 h
Aim of Experiment: To determine trace lead in water by graphite furnace AAS at 283.3 nm.

Theoretical Foundations

Electrothermal (graphite furnace) AAS atomizes the whole microlitre sample in a confined tube, giving ~100× better detection limits than flame AAS (~0.1 µg L⁻¹ for Pb). Programmed drying, ashing and atomization steps separate matrix removal from atomization; the 283.3 nm Pb line is monitored with Zeeman background correction.

Required Apparatus & Chemicals

Laboratory Apparatus

  • GFAAS spectrometer
  • Pb hollow-cathode lamp
  • Autosampler
  • Micropipettes
  • Argon supply

Chemical Reagents

  • Pb standard (100 µg L⁻¹)
  • Nitric acid (ultrapure)
  • Matrix modifier (NH4H2PO4/Mg(NO3)2)
  • Water sample

Step-by-Step Practical Procedure

  1. Acidify sample with ultrapure HNO3.
  2. Prepare Pb standards 0-100 µg L⁻¹.
  3. Program: dry 110 °C, ash 600 °C, atomize 1800 °C.
  4. Inject 20 µL with matrix modifier.
  5. Record peak-area absorbances.
  6. Determine Pb; compare with the 10 µg L⁻¹ guideline.

Observations & Expected Results

Sharp transient peaks; peak area linear with Pb; typical slope ≈ 3.6 A·L mg⁻¹ (peak area).

Worked Stoichiometric Calculations

Problem: Slope 3.6 A·L mg⁻¹ (peak area). A sample gives A = 0.180. Find Pb in mg L⁻¹ and µg L⁻¹.

Working: c = 0.180/3.6 = 0.050 mg L⁻¹ = 50 µg L⁻¹ Pb (above the 10 µg L⁻¹ guideline).

Safety Precautions & Hazard Control

  • Lead is toxic; micropipette carefully.
  • Ultrapure reagents; contamination dominates at µg L⁻¹.
  • Never exceed the tube's temperature program.
  • Argon must flow to protect the graphite tube.

Waste Disposal Protocol

Trace Pb waste to the lead waste bottle.

Viva Voce Oral Defense Questions & Answers

Q: Why is GFAAS ~100× more sensitive than flame AAS?

A: The entire sample is atomized in the light path for seconds, versus milliseconds of dilute flame gases, hugely increasing atom density.

Q: What is the matrix modifier for?

A: It stabilizes Pb during ashing so matrix can be burnt off without losing analyte.

Q: Why peak area rather than height?

A: Area integrates the whole atomization pulse and is less affected by atomization kinetics.